User's Manual

G4-Global System User’s Guide | 49
5. To enter the ngerstick reading taken from your meter, press the TOP or
BOTTOM TOGGLE button until you nd the correct reading and then press
the SELECT button.
a. Pressing the
TOP or BOTTOM TOGGLE button lets you scroll by
1 mg/dL. To scroll faster, hold down the
TOP or BOTTOM TOGGLE button.
6. Con rm that the ngerstick reading you entered is
correct.
a. If the ngerstick reading displayed is correct,
press the
LEFT TOGGLE button to highlight
“OK” and then press the
SELECT button.
b. If the ngerstick reading displayed is incorrect,
press the
SELECT button (with “Cancel”
highlighted) to return to the Enter BG screen. Repeat the steps for
re-entering the correct reading.
c. If you do not press any button, the Receiver will “time out” and no
ngerstick reading will be entered.
7. The Enter BG Processing screen will appear to let
you know the ngerstick reading is being processed
for calibration.
8. For Startup Calibration, repeat these steps for the second ngerstick reading.
9. A glucose reading will appear on the Receiver right away, and glucose
readings will be updated every 5 minutes.
10. If readings do not appear immediately, see Chapter 9, Section 9.2.