User's Manual
G4-Global System User’s Guide | 161
LOW GLUCOSE ALARM: Your System reading is at or below the
factory Low Glucose ALARM level (55 mg/dL). You cannot change
this ALARM level.
RISE ALERTS: Either the Rise Alert or Rapid Rise Alert may appear,
depending on the Rise Alert level you set.
Rise Alert: Your glucose levels are rising at 2 mg/dL per minute or
more.
OR
Rapid Rise Alert: Your glucose levels are rising fast at 3 mg/dL per
minute or more.
FALL ALERTS: Either the Fall Alert or the Rapid Fall Alert may
appear, depending on the Fall Rate Alert level you set.
Fall Alert: Your glucose levels are falling at 2 mg/dL per minute or
more.
OR
Rapid Fall Alert: Your glucose levels are falling fast at 3 mg/dL per
minute or more.
OUT OF RANGE ALERT: The Transmitter and Receiver are not
talking and you will not receive glucose readings.
SENSOR FAILED SCREEN: The Sensor is not working. You will
need to replace the Sensor at this time.
TIME/DATE LOSS ALERT: The Receiver backup battery is drained,
so you need to reset the Receiver time and date. The Time Loss Alert
will automatically take you to the Time/Date Setting Screen.
If you were in the middle of a Sensor Session, this Alert will end your
session. You must recharge the Receiver battery. Then, you must set
the correct date and time and start a new session.