a. Field strengths from fi xed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fi xed RF
transmitters, an electromagnetic site survey should be considered. If the
measure fi eld strength in the location in which the Receiver is used exceeds
the applicable RG compliance level above, the Receiver should be observed
to verify normal operation. If abnormal performance is observed, additional
measures may be necessary such as reorienting or relocating the Receiver.
b. Over the frequency range 150 KHz to 80 MHz, fi eld strengths should be less
than 3 V/m.
11
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Conducted RF
IEC 61000- 4-6
3 Vrms
150 k Hz to
80 MHz
3 V Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Radiated
RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m Field strengths from fi xed RF
transmitters, as determined by
an electromagnetic site survey
a
should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity
of equipment marked with
following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected
by absorption and refl ection from structures, objects and people