User's Manual

G4-Global System User’s Guide | 151
Immunity Test IEC 60601 Test
Level
Receiver
Compliance
Level
Electromagnetic
Environment Guidance
Power Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power Frequency
magnetic elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment
Note: UT is the a.c. main voltage prior to application of the test level
Guidance and Manufacturers Declaration- Electromagnetic Immunity
The Receiver is intended for use in the electromagnetic environment speci ed
below. The customer or the user of the Receiver should assure that it is used in
such an environment.
Electromagnetic Environment Guidance
Portable and mobile RF communications equipment should be used no closer to any
part of the Receiver, including cables than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
Recommended Separation Distance
d =1.2 P
1/2
d = 1.2 P
1/2
80 MHz to 800 MHz
d = 2.3 P
1/2
800 MHz to 2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected
by absorption and re ection from structures, objects and people