User's Manual
150 | Continuous Glucose Monitoring System
Immunity Test IEC 60601 Test
Level
Receiver
Compliance
Level
Electromagnetic
Environment Guidance
Electromagnetic
Environment
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood,
concrete or ceramic tile.
If fl oors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical Fast
Transient /burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Main power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC 61000-4-5
± 1 kV
differential mode
± 2 kV
common mode
± 1 kV
differential mode
± 2 kV
common mode
Main power quality
should be that of a typical
commercial or hospital
environment
Voltage
Dips, Short
Interruptions and
Voltage Variations
on Power Supply
Input Lines
IEC 61000-4-11
< 5 % UT
(>95 % dip in
UT) for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
< 5% UT
(>95% dip in UT)
for 5 sec
< 5 % UT
(>95 % dip in
UT) for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
< 5% UT
(>95% dip in UT)
for 5 sec
Main power quality
should be that of a typical
commercial or hospital
environment
11
Guidance and Manufacturer’s Declaration- Electromagnetic Immunity
The Receiver is intended for use in the electromagnetic environment specifi ed
below. The customer or the user of the Receiver should assure that it is used in
such an environment.