User's Manual
G4-Global System User’s Guide | 123
For Single-Patient Use
Cleaning
1. Use an alcohol wipe, or equivalent to clean the outside of the Receiver and
Transmitter. Avoid using wipes that contain adhesive properties. (e.g. Smith +
Nephew IV Prep
™
).
2. Place the Receiver and Transmitter on a clean, dry cloth and air dry for 2-3
minutes.
For Multi-Patient Use
Preparation
1. Use precaution when handling products worn by another person.
2. Wash hands thoroughly.
3. Wear personal protective equipment as appropriate (gloves, protective
goggles, gowns, etc.).
Disinfection
4. Wipe down the outer surface of each component with Steris
®
Coverage
®
Spray TB Plus, or equivalent sterilant.
5. Allow contact time for 3 minutes at room temperature (20°C/68°F).
NOTE:
• Always clean the Receiver and Transmitter after each use.
• Do NOT use bleach or acetone to clean the Receiver and Transmitter.
NOTE: Do NOT get the DexCom Receiver wet. Ensure the wipe is slightly
damp, but not wet enough to leave drops of liquid on the Receiver surface
(particularly the openings and buttons).