User's Manual

G4-Global System User’s Guide | 111
9.1 SENSOR INSERTION TROUBLESHOOTING
Sensor Insertion Dif culties
I am having trouble taking out the Safety Lock/Transmitter Key:
- Make sure to pull straight out using the arrows on the Safety Lock as a guide.
- Do not wiggle back and forth.
I am not able to pull the Collar up:
- Make sure the plunger is completely pressed down before pulling the
collar up.
- When pulling the Collar up you should hear 2 clicks.
- Try using more force when pulling the collar up.
I am not able to remove the Applicator from the Sensor Pod:
- Pull the Collar all the way up. It should be very close
to the top of the Applicator.
- Make sure the Transmitter Latch is down before
squeezing the Release Tabs.
- Then squeeze the ribbed Release Tabs (on the
side), and lift the Applicator away from your body.
Removing the Transmitter Latch is dif cult:
- Hold the Sensor Pod down with one hand and twist the Latch with the other
hand to remove it.
- Do not try to snap it straight off.
Sensor Pod is Not Sticking Long Enough
Do not use any cream or lotion on your skin where you attach the Sensor Pod.
Clean the skin and make sure it is completely dry before you attach the Sensor
Pod. Do not leave any substance on the skin where the Needle inserts.
You may use medical tape (such as Blenderm
TM
) over the white adhesive patch
of the Sensor Pod, but do not place the tape over the Transmitter or the plastic
parts of the Sensor Pod.
The G4-Global Continuous Glucose Monitoring System is a glucose-
monitoring device indicated for detecting trends and tracking patterns in
adults (age 18 and older) with diabetes.
Release the Applicator