DeVilbiss Polino-Aide® Coinpressor /Nebulizer Service Manual Models 56500 & 5650C Putmo,4/de Printed in the U.S.A. @1996 DHC, Inc. LT-1469 Rev.
Purpose To provide a portable means of supplying compressed air to devices which do not require a large volume of air, such as nebulizers, atomizers, and syringes. The Installation Lower access door to expose the air outlet connector. The compressor is set up for use with slip-on hose connectors.Attach the Maintenance The air filter (14) should be changed whenever it turns completely gray in color. Remove the filter cap (15) by pulling outward.
Important Parts of the Pulmo-Aide 0 @ @ 0 18 _, "'=' 2 = 0 9 8 0 5 7 � 6 I I 10 I REPLACEMENT PARTS 1. 5650D-606 2. 5610D-609 3. 5610D-614 4. 5. 5610D-617 5610D-614 6. 5650D-613 7. 5610D-614 8. 5610D-614 9. 5650D-604* 10. 5650D-615 11. 5650D-610 5650H-610 12. 5650D-614 13. 5650D-609 14. 5650D-602 5650D-601 15. 5650D-612 16. 5650D-603* 17. 5650D-605* 18.
Troubleshooting SYMPTOMS POSSIBLE CAUSES REMEDIES A. Reduced air output. 1. Dirty air inlet filter. 2. Worn diaphragm or valve plate. 3. Air leak. 1. Replace filter. 2. Replace diaphragm or valve plate. 3. Tighten hose connections and tighten head screws. B. Compressor labors when air device is not being used. 1. Occluded nebulizer or tubing. 1. Install some means of venting excess air when device is not in use or use PIN 656 pressure regulator.