Instruction Manual

EN - 11
SPECIFICATIONS
DEVILBISS 5
-
LITER SERIES
Catalog Number 525DS, 525DS-Q 525KS, 525KS-LT 525PS
Delivery Rate (Lower delivery rates available
or low fow applications)***
0 5 to 5 LPM 0 5 to 5 LPM 0 5 to 5 LPM
Maximum Recommended Flow (@ nominal
outlet pressures o zero & 7 kPa)**
5 LPM 5 LPM 5 LPM
Outlet Pressure 8.5 ± 0.5 psig (58.6 ± 3.5 kPa) 8.5 ± 0.5 psig (58.6 ± 3.5 kPa) 8.5 ± 0.5 psig (58.6 ± 3.5 kPa)
Auxiliary Oxygen Port **
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
Electrical Rating 115 V, 60 Hz, 3 3 Amp
220-230 V~, 50 Hz, 1 55 Amp
230 V~, 60 Hz, 1 9 Amp
220-230 V~, 60 Hz, 1 68 Amp
Operating Voltage Range 97-127 V~, 60 Hz
187-253 V~, 50 Hz
195-253 V~, 60 Hz
187-253 V~, 60 Hz
Oxygen Percentage 1-5 LPM=87%-96% 1-5 LPM=93%±3% 1-5 LPM=93%±3%
Operating Atmospheric Pressure
1010 hPa to 840 hPa
0-1500 M (0-4921 t)
Across the voltage range:
No degradation of performance
Across the voltage range:
No degradation of performance
Across the voltage range:
No degradation of performance
840 hPa to 616 hPa
1500-4000 M (4921-13123 t)
Tested at nominal voltage only:
No degradation of performance
Tested at 230V/50Hz only:
No degradation of performance
Tested at 230V/60Hz only:
No degradation of performance
Operating Temperature Range 41˚F (5˚C) to 95˚F (35˚C) 41˚F (5˚C) to 95˚F (35˚C) 41˚F (5˚C) to 95˚F (35˚C)
Operating Relative Humidity Range 15% to 93%, non-condensing 15% to 93%, non-condensing 15% to 93%, non-condensing
Power Consumption
310 Watts Average
275 Watts @ 1.2 LPM & below
230V / 50 Hz - 312 Watts Average
230V / 50 Hz - 296 Watts Average @
1 2 LPM & below
230V / 60 Hz - 387 Watts Average
230V / 60 Hz - 369 Watts Average @
1 2 LPM & below
230V / 60 Hz - 334 Watts Average
230V / 60 Hz - 297 Watts Average @
1 2 LPM & below
Weight 36 lbs. (16.3 Kilograms) 36 lbs. (16.3 Kilograms) 36 lbs. (16.3 Kilograms)
Safe Working Load 53 lbs. (24 Kilograms) 53 lbs. (24 Kilograms) 53 lbs. (24 Kilograms)
Sound Pressure Level at 3 and 5 LPM
(ISO 80601-2-69)
525DS
50.9 dBA @ 3 LPM
50.7 dBA @ 5 LPM
525DS-Q
46.7 dBA @ 3 LPM
46.7 dBA @ 5 LPM
525KS
47.9 dBA @ 3 LPM
47.9 dBA @ 5 LPM
525KS-LT
49.6 dBA @ 3 LPM
49.4 dBA @ 5 LPM
45.4 dBA @ 3 LPM
45.3 dBA @ 5 LPM
Sound Power Level at 3 and 5 LPM
(ISO 80601-2-69)
525DS
54.7 dBA @ 3 LPM
54.5 dBA @ 5 LPM
525DS-Q
50.4 dBA @ 3 LPM
50.4 dBA @ 5 LPM
525KS
51.6 dBA @ 3 LPM
51.7 dBA @ 5 LPM
525KS-LT
53.4 dBA @ 3 LPM
53.2 dBA @ 5 LPM
49.2 dBA @ 3 LPM
49.1 dBA @ 5 LPM
Sound Level (ISO 8359:1996)
48 dBA (525DS)
46 dBA (525DS-Q)
40 dBA (50 Hz (525KS)
48 dBA (50 Hz (525KS-LT)
Alarm Sound Level > = 62 dBA > = 62 dBA > = 62 dBA
Dimensions
24 5”H x 13 5”W x 12”D
(62.2 x 34.2 x 30.4 cm)
24 5”H x 13 5”W x 12”D
(62.2 x 34.2 x 30.4 cm)
24 5”H x 13 5”W x 12”D
(62.2 x 34.2 x 30.4 cm)
Maximum Limited Pressure
Normal Condition: 9 PSIG (62.0 kPa),
Single Fault Condition: 27.6 PSIG (190.3 kPa)
Normal Condition: 9 PSIG (62.0 kPa),
Single Fault Condition: 27.6 PSIG (190.3 kPa)
Normal Condition: 9 PSIG (62.0 kPa),
Single Fault Condition: 27.6 PSIG (190.3 kPa)
Operating System Time Cycle / Pressure Swing Time Cycle / Pressure Swing Time Cycle / Pressure Swing
Low Oxygen Indicator
<82% low oxygen
<60% very low oxygen
<82% low oxygen
<60% very low oxygen
<82% low oxygen
<60% very low oxygen
Storage Conditions -13˚F (-25˚C) to 158˚F (70˚C), humidity range
o 15% to 93% non-condensing
-13˚F (-25˚C) to 158˚F (70˚C), humidity range
o 15% to 93% non-condensing
-13˚F (-25˚C) to 158˚F (70˚C), humidity range
o 15% to 93% non-condensing
Equipment Class and Type
Class II Equipment Double Insulated;
Type B Applied Part, IP21
Class II Equipment Double Insulated;
Type B Applied Part, IP21
Class II Equipment Double Insulated;
Type B Applied Part, IP21
Approval Body and Safety Standard
TUV
ANSI/AAMI ES60601-1:2005+A2 (R2012) +A1
IEC 60601-1-6:2010
IEC 60601-1-11:2015
ISO 80601-2-69:2014 ***
CAN/CSA-C22 2 No 60601-1:14
CAN/CSA-C22 2 No 60601-1-6:11
CAN/CSA-C22 2 No 60601-1-11:15
CAN/CSA-C22 2 No 80601-2-69:16
TUV approved for 50 Hz only to
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
TUV
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
CE mark No Yes Yes
EMC Compliance To EN60601-1-2 EN60601-1-2 EN60601-1-2
**
CAUTION The maximum recommended fow is 3 LPM when an oxygen bottle is being lled with oxygen rom the auxiliary oxygen port.
***
NOTE Use o the 515LF-607 low output fow meter package or other low output fow meter accessory will prevent the device rom meeting the requirements o ISO-80601-2-69:2014
Section 201 13 2 101
Specications subject to change without notice.
A-5 25D S