Product Manual

0044-0391-0M Rev A y solo Digital Clinical Scale
21
DISPLAY MESSAGES
The solo Digital Clinical Scale is equipped with software that tests various portions of the
scale's circuitry and verifies proper operation. Should a problem be detected, a message will
be displayed alerting the operator to that condition. The following lists the messages
displayed and their meaning:
DISPLAY
MEANING
99
This will be displayed for a few seconds when the scale cannot establish
zero
and then turn the scale off. Refer to the TROUBLESHOOTING
section.
!6;`
Displayed when the batteries are near the point they need to be replaced.
#$
Weight on scale exceeds 570 lb (260 kg).
TROUBLESHOOTING
Your scale has been designed to provide you with years of trouble-free operation. In spite of
this, troubles sometimes happen. Before calling for service assistance you should make
some initial checks to verify that a problem does exist. The following describes several types
of symptoms along with suggested remedies.
PROBLEM
POSSIBLE SOLUTIONS
Display does not
turn on
Battery Operation:
y Check if batteries are installed and correctly.
y Are batteries discharged? Remove old batteries and re
place with
new ones.
AC Operation:
y Is 9V AC/DC power adapter fully inserted into wall receptacle?
y
Check wall receptacle for proper AC power. Try another electrical
appliance in same receptacle, does it work?
y Check circuit breaker.
y
Has there been power failure?
Incorrect weight
displayed
Ensure that the scale platform isn't touching an adjacent object.
Have proper operation procedures been followed?
Display shows
99
and shuts
off
If platform is not empty (patient is on scale) when powering up, display
will
show 99 and turn off.
Remove any weight from platform (have
patient
step off scale) and power up the scale again.
If scale continues
to show
99
and turn off, consult your scale service representative.
Display shows a
moving small
6
If the
display shows a moving small
6
when powering up, insure the
scale is stable and wait for a moment until the display shows
..
If scale continues to show a moving small
6
, please
consult your scale
service representative.