17
SETUP, CONT.
"0f (Months)
The display will show 00f, the current Months setting. . If value displayed is
acceptable, press the
key to save it and proceed to the next prompt. Otherwise,
press the ©
©
or ª keys to select a new value from 01 to 12 and then press the key
to save it and proceed to the next prompt. Note that 01 represents January, 02
represents February so on.
NOTE: The default month setting is 01 (January).
3f (Day)
The display will show 33f, the current Day setting. . If value displayed is
acceptable, press the
key to save it and proceed to the next prompt. Otherwise,
press the
©
or ª keys to select a new value from 01 to 31 and then press the
key to save it and proceed to the next prompt.
NOTE: The default day setting is 01 (first day of month).
* (Automatic Shutoff)
The display will show 1:,AA, the current Automatic Shutoff setting. . If value
displayed is acceptable, press the
key to save it and proceed to the next prompt.
Otherwise, press the
©
or ª keys to select a new value and then press the key
to save it and proceed to the next prompt. Allowable values are 00 to 09 minutes in 1
minute increments. Note that selecting 00 disables the automatic shutoff feature.
NOTE: The default Automatic Shutoff setting is * (3 minutes).
2<gg (Buzzer Setting)
In operation mode, if an incorrect key is pressed, the scale will sound two beeps as a
warning and the operation will not be processed.
The display will show 2<gg, the current Buzzer setting. If value displayed is
acceptable, press the
key to save it and proceed to the next prompt. Otherwise,
press the
©
or ª keys to select enable or disable the Buzzer and then press the
key to save it and proceed to the next prompt. Allowable values are 0 (disable) or 1
(enable).
NOTE: The default Buzzer setting is 1 (enable).
Setup Complete
Once the key has been pressed at the 2<gg prompt, all Setup settings (current or new
values) will be saved and the scale will turn off.