Owner`s manual

4
SETUP AND CALIBRATION, CONT.
UNLOAD WEIGHT
With the display showing uLduLd, remove the 2 lb test weight from the scale tray. Allow
approximately five seconds for the scale to stabilize, then press the ON/TARE key to continue
the calibration. Dashes will be displayed from the left to the right of the display momentarily,
then the display will change to untunt.
If ErAErA is displayed during the LOdLOd or ulduld calibration sequence, remove the test weight
from the scale and verify that it is the correct weight (2 lbs) or insure that the scale is
empty and nothing is touching the scale tray. The display will return to LOdLOd or ulduld
allowing you to continue with calibration and setup.
WEIGHING UNITS
With the display showing untunt, press the WT/UNIT key to display the current setting for the
default weighing unit. The display will show ------ and turn on the red indicator for the weight
unit (oz or gm) or 11 and the red indicator for (gm) for the "Grams only" mode. Press the
WT/UNIT key until the desired weight unit is indicated. Press the ON/TARE key to save the
setting.
AUTOMATIC SHUTOFF
The display will show ASHASH. Press the WT/UNIT key to display the current setting. Press the
WT/UNIT key until the desired number of minutes of inactivity before the scale will
automatically shutoff is displayed. Values of 1, 2, 5, 10, or 15 minutes or OFF are available.
Selecting OFF will disable the automatic shutoff feature. Press the ON / TARE key to save the
setting.
POWER UP ZERO
The display will show PuOPuO. Press the WT/UNIT key to display the current setting. Press the
WT/UNIT key to select YESYES to enable or no no to disable. Press the ON / TARE key to save the
setting.
DONE
The scale will briefly display dnEdnE, then reset to the Operation weight mode.
SETUP REVIEW
The Setup Review mode allows you to review and change several operational parameters as
desired without having to enter the setup and calibration mode. To enter the setup review mode,
you need to turn the scale off by pressing the WT/UNIT and ON-TARE keys simultaneously and
then pressing and holding the ON-TARE key until the red indicator of the selected weight units
begins to flash. After the red indicator begins to flash, release the ON-TARE key. The display will
show untunt, the prompt for the weighing units. Refer to the instructions listed in the Setup and
Calibration section for information on how to change these parameters. The parameters in the
setup review that can be change are as follows:
WEIGHING UNITS AUTOMATIC SHUTOFF
Location of
Calibration
Switch