Owner`s manual

3
DISPLAY CODES
CALIBRATION REQUIRED - Indicates improper stored calibration data, calibration
is necessary.
OVER CAPACITY - Indicates a weight exceeding the capacity of scale has been
placed on the scale tray.
ERROR - NOV-RAM test failure.
ANALOG FAILURE - This could mean a damaged or defective load cell, a failed
component in the analog circuity of the scale or the platform isn't on or is on
incorrectly.
SETUP AND CALIBRATION
Your scale has been thoroughly tested and calibrated before being shipped to you and should not
require adjustment. In the event the scale should need re-calibration, the following calibration
procedure should be performed. To maintain the scale's high degree of accuracy, it is
recommended that a qualified technician with certified test weights re-calibrate the scale.
In order to perform Setup and Calibration, it is necessary to remove the "Do Not Remove" sticker
from the lower right side of the scale. With the sticker removed, locate the Setup and Calibration
switch in the opening closest to the front of the scale. Refer to the figure on the next page.
NOTE: If operating from batteries, install the batteries making certain that they are
positioned in accordance with the drawing on the circuit board. If using the
optional AC power adapter, insert the plug into the jack on the back of the scale.
BEGIN SETUP and CALIBRATION
1. Making certain the scale tray is empty, press and hold the setup and calibration switch located
on the lower right side of the scale. Refer to the figure on the next page.
2. Press the ON/TARE key to turn the scale on. Release the setup and calibration switch when
the display is showing SCASCA.
SCALE MODEL
The display will show SCASCA. Press the WT/UNIT key to display the current setting. For a PS-5
the display should show PP--55. Press the WT/UNIT key until PP--55 is displayed. Press the
ON/TARE key to save the setting.
CALIBRATION
The display will show CALCAL. Press the WT/UNIT key to display the current setting. To calibrate
the scale, press the WT/UNIT key until YESYES is displayed, then press the ON/TARE key. The
display will change to LOdLOd.
To skip calibration, press the WT/UNIT key until nono is displayed, then press the ON/TARE key
to continue the setup.
LOAD WEIGHT
With the display will showing LOdLOd, place a (2) lb certified test weight on the scale tray. Allow
approximately five seconds for the scale to stabilize. Press the ON/TARE key to start the
calibration. Dashes will be displayed from the left to the right of the display momentarily, then
the display will change to uLduLd.
OCOC
CALCAL
ErrErr
ErAErA