Product Manual
<LF>A<CR> Scale will respond with:
<LF>MedVue<CR>
<LF>XH<CR> Scale will respond with the current height:
<LF>Height<CR>
<LF>XB<CR> Scale will respond with the current BMI:
<LF>BMI<CR>
<LF>XI<CR> Scale will respond with the currently entered ID:
<LF>ID<CR>
<LF>XA<CR> Scale will respond with current weight, height, BMI, and ID in comma delimited
format:
<LF> weight (see <LF>W<CR> above), Height, BMI, ID<CR>
If the MedVue weight analyzer has been set to continuous mode of outputting weight, it will
output the weight in the format listed above for <LF>W<CR> every 500 milliseconds.
MVWIFI RESET
If you are having issues connecting to the MVWIFI you may want to try and reset the MVWIFI
option card.
With the weight analyzer on and in the Gross weight mode:
1. Press and hold the CLEAR ( ) key for approximately 3 seconds.
2. The display will change to show SETUP REVIEW.
3. Release the CLEAR ( ) key.
4. The display changes to show SETUP.
5. Press the ID/HEIGHT key until the ETHERNET/WIFI prompt is displayed.
6. Press the ENTER key until the RESET ETHERNET prompt is displayed.
7. Change this setting to YES (press the 1/YES key) and then press the ENTER key.
8. Wait for the MedVue to reset the MVWIFI.
9. Refer to the Network Configuration section of this manual to setup network connection.
8555-M519-O1 MedVue MVWIFI Operation
11