<LF>W<CR> String sent to the MedVue weight analyzer will respond with the SMA weight
string:
<LF>S1GM^DDDDDDDDDDUUU<CR>
Where:
<LF> = Line feed character (hex 0A)
S = Status (‘O’ = Over Cap, ‘Z’ = Center Zero, ‘U’ = Below Zero, ‘E’ = Error)
1 = The number ‘1’
G = Mode of operation (‘G’ = Gross, ‘N’ = Net, ‘T’ = Tare)
M = Motion bit (‘M’ = Motion, ‘ ‘ = Settled)
^ = Space
DDDDDDDDDD = Weight with decimal point if necessary
UUU = Units (e.g. ‘lb ‘, ‘kg ‘, ‘ton’, etc..)
<CR> = Carriage return (hex 0D)
<LF>P<CR> Weight analyzer will respond with all information formatted with each item on its
own line.
<LF>J<CR> Weight analyzer will respond with all information formatted with each item
separated by commas (comma delimited) followed by a carriage return.
<LF>Z<CR> Zero the scale within the constraints of the zero settings.
<LF>T<CR> Scale attempts to tare itself setting the tare weight to the current gross weight.
<LF>T<xxxxxx.xxx><CR> Scale attempts to set the weight (represented as<xxxxxx.xxx> in
the string) to the requested value.
<LF>M<CR> Scale returns the current tare weight.
<LF>U<CR> Scale will toggle between calibration units if conversion units have been set up
in scale calibration.