Model MV1 Medical Weight Analyzer Owner’s Manual 8555-M512-O1 Rev A 03/13 1
8555-M512-O1 Rev A 03/13
INTRODUCTION Thank you for purchasing our MedVue® Model MV1 Medical Weight Analyzer. It has been manufactured with quality and reliability at our factory in Webb City, MO USA. The MV1 has been tested before leaving our factory to insure accuracy and dependability for years to come. This manual is provided to guide you through installation, operation and maintenance of the MedVue Model MV1 Medical Weight Analyzer.
FCC COMPLIANCE STATEMENT This equipment generates uses and can radiate radio frequency and if not installed and used in accordance with the instruction manual, may cause interference to radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC rules, which are designed to provide reasonable protection against such interference when operated in a commercial environment.
TABLE OF CONTENTS SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Optional Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SITE PREPARATION REQUIREMENTS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . .
IV 8555-M512-O1 Rev A 03/13
SPECIFICATIONS Power Requirements . . . . . . . . 6 “AA” cell Alkaline, Ni-Cad or NiMH batteries (not included) OR an optional 12 VDC 1.25A wall plug-in AC power adapter (Cardinal part number MV1PWR). Operating Temperature . . . . . . 14 to 104 ºF (-10 to +40 ºC) Display . . . . . . . . . . . . . . . . . . . Weight: 5 digit, seven segment, 7/8” (22.23 mm) high LCD Height: 4 digit, seven segment, 1/2" (12.7 mm) high LCD BMI: 3 digit, seven segment, 1/2" (12.7 mm) high LCD ID: 14 digit, 14 segment, .35” (8.
SITE PREPARATION REQUIREMENTS The MedVue Model MV1 Medical Weight Analyzer is a precision weight indicating instrument. As with any precision instrument, it requires an acceptable environment to operate at peak performance and reliability. This section is provided to assist you in obtaining such an environment. Environmental The MedVue Model MV1 Medical Weight Analyzer meets or exceeds all certification requirements within a temperature range of 14 to 104 °F (-10 to +40 °C).
INSTALLATION Unpacking Carefully remove the MedVue Model MV1 Medical Weight Analyzer from the shipping carton and inspect it for any damage that may have taken place during shipment. You should keep the carton and packing material for return shipment if it should become necessary. The purchaser is responsible for filing all claims for any damages or loss incurred during transit.
Mounting The MedVue Model MV1 Medical Weight Analyzer is mounted to the scale (or an optional desktop/wall bracket) using a swivel ball and socket assembly. To install the MV1 Rubber Ball in to the Swivel Ball Socket, slightly loosen the Adjustment Knob to allow the two parts of the socket to open enough to insert the Rubber Ball. Adjust the MV1 to the desired viewing angle and direction and then tighten the Adjustment Knob. Adjustment Knob Loosen slightly to insert Rubber Ball in to Swivel Ball Socket.
INTERCONNECTIONS All input, output, power and option connections to the MV1 are made on the its rear panel. Connections for the Load Cell input and the Serial I/0 are made using 9-pin "D" shaped connectors. The Height Rod option (Detecto DHR, Digital Height Rod) and Ethernet (MVIP) options use snap-in modular connectors. The USB port is a device (or upstream) port using an industry standard “B” connector. The optional 12VDC wall plug-in UL/CSA listed AC power adapter is connected using a power jack.
Serial I/O Port The serial I/O port is a 9-pin "D" connector on the rear panel of the MV1. Figure No. 5 shows the Serial I/0 connector along with the identity of the pins used. Make certain that the pins are correctly identified before soldering a wire to them. Use the connector retaining screws to hold the serial cable connector securely to the rear panel. PIN NO. 2 3 5 9 FUNCTION DATA INPUT (RXD) DATA OUTPUT (TXD) SIGNAL GROUND (GND) VCC or 5VDC 5 4 9 Note that pins 1, 4 and 6 through 8 are not used.
Battery Status Symbol A battery status symbol will be displayed when batteries are installed. The number of bars indicates the level of charge remaining in the batteries. 4 Bars = 100% 3 Bars = 75% 2 Bars = 50% 1 Bar = 25% No Bars = Low Battery Low Battery Indicator When the batteries are near the point they need to be replaced (Alkaline) or recharged (NiCad or NiMH), the battery status symbol will not show any bars.
KEYPAD FUNCTIONS DO NOT operate the keypad with pointed objects (pencils, pens, etc). Damage to keypad resulting from this practice is NOT covered under warranty. Figure No. 7 This is the ON/OFF key. With the MV1 off, pressing this key will apply power to it and turn on the display. If the MV1 is already on, pressing the key will turn it off. Basic Functions This is the ZERO key. It is used to reset the display to zero up to the selected limit of either 4% or 100% of the scale capacity.
Basic Functions, Cont. This is the LOCK/RELEASE key. It is used to lock and unlock the display. If the HOLD feature was enabled during setup and calibration, pressing this key (after obtaining a stable weight) will cause the MV1 to lock onto the weight. Pressing this key again will unlock the display and return it to zero. If the AUTO LOCK feature was enabled, the weight display will lock after placing a load on the scale and obtaining a stable weight value.
Body Mass Index (BMI) Functions This is the CLEAR (BACKSPACE) key. It is used to clear an incorrect entry from the display without processing the data. If an incorrect entry is made, press this key and re-enter the correct data. When entering the ID, pressing this key will act as a “backspace” key (it will back up one space and delete the character in that space). NOTE: The CLEAR key must be pressed before the PRINT/ENTER key to ensure the data is not processed. This is the PRINT/ENTER key.
ANNUNCIATORS The annunciators are turned on to indicate that the display is in the mode corresponding to the annunciator label or that the status indicated by the label is active. Weight Display Battery Status Height or Clock Display BMI Display Displayed when optional hardware is installed. Messages Display Figure No. 8 GROSS The GROSS annunciator is turned on to show that the weight displayed is the gross weight. Gross weight will be displayed when no tare weight value is stored.
CAL The CAL annunciator will be turned on after pressing the calibration switch, to show that the MV1 is in the Setup and Calibration mode. lb This annunciator is turned on to indicate that the displayed weight is in pounds. kg The kg annunciator is turned on to indicate that the displayed weight is in kilograms. ' '' The ( ' ) feet and ( '' ) inches annunciators are turned on when the displayed height measurement is in feet and inches.
OPERATION Basic Operation To Weigh 1. Press ON/OFF key to turn the MV1 on. 2. Press ZERO key to zero weight display. ZERO, GROSS and lb or kg annunciator will turn on to show that the scale is ready for use. 3. Place patient on scale and read weight display. 4. If a printer is connected to scale, press PRINT/ENTER key to print a ticket. 5. Remove patient from scale. Zero Weight Display 1. In Gross Weight mode (GROSS annunciator on), press ZERO key. 2. Weight display will return to zero.
Tare Weight Entry (Tare weight is known) 1. In Gross Weight mode (GROSS annunciator on), press the TARE key. 2. The Display will change to show ENTER TARE WT and the TARE annunciator will turn on. 3. Use the numeric keys and enter a tare weight. Note that when entering tare weights, a maximum of 4 digits can be entered and the numbers advance from right to left in the display.
Tare Storage and Recall NOTE: ID must be enabled to use stored Tare. Refer to SYSTEM Menu, Enable ID. To Store Tare Weight 1. In Gross Weight mode (GROSS annunciator on), press the TARE key. 2. The Display will change to show ENTER TARE WT and the TARE annunciator will turn on. 3. Use the numeric keys and enter a tare weight. Note that when entering tare weights, a maximum of 4 digits can be entered and the numbers advance from right to left in the display.
Basic Operation with ID To Weigh 1. Press ON/OFF key to turn the MV1 on. 2. Press ZERO key to zero weight display. ZERO, GROSS and lb or kg annunciator will turn on to show scale is ready for use. 3. Press ID/HEIGHT key. 4. Display will change to show Id. 5. Using numeric keys, enter up to an 14 digit alphanumeric identification number. 6. Press PRINT/ENTER key. 7. Place patient on scale 8. Read weight display. 9. If a printer is connected to scale, press PRINT/ENTER key to print a ticket. 10.
Tare Weight Entry (Tare weight is known) 1. In Gross Weight mode (GROSS annunciator on), press TARE key. 2. Display will show )0 or previously entered tare and TARE annunciator will turn on. 3. If weight displayed is acceptable, press PRINT/ENTER key to save it. Otherwise, use the numeric keys and enter a new tare weight. Note that when entering tare weights, a maximum of 4 digits can be entered and the numbers advance from right to left in the display.
Body Mass Index (BMI) Operation Using Mechanical Height Rod or Manually Entering Height 1. Press ON/OFF key to turn the MV1 on. 2. Press ZERO key to zero weight display. ZERO, GROSS and lb or kg annunciator will turn on to show scale is ready for use. 3. Perform tare operation if required. Refer to Push Button Tare or Tare Weight Entry in previous section for instructions on using tare. 4. Press ID/HEIGHT key. 5.
Body Mass Index (BMI) Operation with DHR Using Digital Height Rod IMPORTANT! Before performing the operations on this page, the digital height rod must be in its “starting position” (inner sliding tube must be down completely inside outer stationary tube and headpiece folded flat against stationary tube) prior to turning on the MV1. 1. Make sure height rod is in starting position. 2. Press ON/OFF key to turn MV1 on. 3. Press ZERO key to zero weight display.
Body Mass Index (BMI) Operation with ID and DHR Using Digital Height Rod IMPORTANT! Before performing the operations on this page, the digital height rod must be in its “starting position” (inner sliding tube must be down completely inside outer stationary tube and headpiece folded flat against stationary tube) prior to turning on the MV1. 1. Make sure height rod is in starting position. 2. Press ON/OFF key to turn MV1 on. 3. Press ZERO key to zero weight display.
CALIBRATION SETUP (Protected Setup Parameters) Your MedVue Model MV1 Medical Weight Analyzer has been thoroughly tested and calibrated before being shipped to you. If you received it with a scale, calibration is not necessary. If the MV1 is being connected to a scale for the first time or recalibration is necessary for other reasons, the following describes the steps to perform calibration setup on the MV1. To Enter Calibration Setup: 1.
Navigation Keys, Cont. ID/HEIGHT – (Next Menu or Parameter) If at the menu level, this will take you to the previous menu item in the list below. If SCALE is currently displayed and the setup has been accessed through the calibration switch, then this will return you to TIME SETTINGS. If setup has been accessed through setup review, then pressing this when SYSTEM is displayed will return you to TIME SETTINGS. If at the parameter level the exact item this takes you to may depend on a [parameter condition].
SCALE With SCALE displayed, press the PRINT/ENTER key. The display will change to CAPACITY. Proceed to the CAPACITY (Scale Capacity) parameter. Otherwise, press the ID/HEIGHT key to advance to the ZERO menu. CAPACITY (Scale Capacity) The display will show CAPACITY and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys, enter a new value and then press the PRINT/ENTER key to save it. Allowable values are 1 through 65,535.
SAMPLE RATE (Sample Rate) The display will show SAMPLE RATE and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys, enter a new value and then press the PRINT/ENTER key to save it. The sample rate may be set from a minimum of 1 sample per second to a maximum of 100 samples per second in one sample per second intervals. Allowable values are 1 through 100.
ZERO With ZERO displayed, press the PRINT/ENTER key. The display will change to ZERO LIMIT. Proceed to the Two Percent Zero Tracking Range Limit parameter. Otherwise, press the ID/HEIGHT key to advance to the FILTER menu. ZERO LIMIT (Two Percent Zero Tracking Range Limit) This setting limits the scale to zero only + or - 2% of the scale capacity. If enabled, (ZERO LIMIT = YES) zero tracking and push button zero is limited to 2% of scale capacity.
FILTER With FILTER displayed, press the PRINT/ENTER key. The display will change to FILTER MOdE. Proceed to the Filter Type parameter. Otherwise, press the ID/HEIGHT key to advance to the CALIBRATION menu. FILTER MOdE (Filter Type) The display will show FILTER MOdE and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys, enter a new value and then press the PRINT/ENTER key to save it. Allowable values are 0, 1 or 2.
CALIbRATION With CALIbRATION displayed, press the PRINT/ENTER key. The display will change to CALIbRATE. Proceed to the Perform Calibration parameter. Otherwise, press the ID/HEIGHT key to advance to the GRAVITY menu. CALIbRATE (Perform Calibration) The display will show CALIbRATE and the current setting (NO). If the scale has been previously calibrated and you wish to skip calibration (and retain the internal calibration factor), press the PRINT/ENTER key. The display will change to GRAVITY ENAbLE.
Single Calibration Point LOAdEd CAL WT (Load Calibration Weight) The display will show LOAdEd CAL WT (which is a prompt for entry of the calibration weight value) and placement of this amount of test weight on scale platform. 1. Make certain scale platform is empty and free of debris. 2. When the display is showing LOAdEd CAL WT, the current setting for the calibration weight will be displayed also. 3.
Multi-point Calibration CAL WEIGHT 1 (First Calibration Weight) The display will show CAL WEIGHT 1 (which is a prompt for entry of the first calibration weight value) and placement of this amount of test weight on scale platform. Note that the MV1 defaults to half (½) of the scale capacity. 1. Make certain scale platform is empty and free of debris. 2. When the display is showing CAL WEIGHT 1, the current setting for the first calibration weight will be displayed also.
GRAVITY With GRAVITY displayed, press the PRINT/ENTER key. The display will change to GRAVITY ENAbLE. Proceed to the Enable Gravity Constant parameter. Otherwise, press the ID/HEIGHT key to advance to the LOCK menu. GRAVITY ENAbLE (Enable Gravity Constant) The Detecto MedVue® Model MV1 Weight Analyzer is equipped with a gravity constant function which means that it can be calibrated in one location and then adjusted to match the gravity at the location where it will used.
LOCK With LOCK displayed, press the PRINT/ENTER key. The display will change to HOLd MOdE. Proceed to the Hold Mode parameter. Otherwise, press the ID/HEIGHT key to advance to the SYSTEM menu. HOLd MOdE (Hold Mode) The Hold Mode Sets the locking functionality of the MV1. The display will show HOLd MOdE and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys, enter a new value and then press the PRINT/ENTER key to save it.
32 8555-M512-O1 Rev A 03/13
CALIBRATION SETUP (Non- Protected Setup Parameters) SYSTEM With SYSTEM displayed, press the PRINT/ENTER key. The display will change to LANGUAGE. Proceed to the Language parameter. Otherwise, press the ID/HEIGHT key to advance to the CALIBRATION menu. LANGUAGE (Language) The display will show LANGUAGE and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys, enter a new value and then press the PRINT/ENTER key to save it.
SLEEP MOdE (Sleep Mode) The Sleep Mode feature also conserves power when the MV1 remains unused for a selected period of time (approximately 1 to 9 minutes). When enabled the display will show the date and time and the backlight will be turned off. The Sleep Mode feature requires the MV1 to remain at a stable zero weight to activate. Weight placed on the scale or pressing any key will activate the MV1 and return it to weight mode. The display will show SLEEP MOdE and the current setting.
bARCOdE (Use ID Scanner) This setting determines whether the MV1 allows a scanner to be used to input an ID number to be entered prior to beginning the weighing operation. This applies to both Basic Operation and with the BMI feature enabled. The display will show bARCOdE and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise use the numeric keys 0/NO or 1/YES and enter a new setting. Press the PRINT/ENTER key to save it.
HI RESOLUTION (High Resolution Mode) The High Resolution Mode feature will display the weight at 10 times the standard resolution. For example, a MV1 configured for 400 lb x 0.2 lb will display weight in high resolution mode at 400 lb x 0.02 lb. NOTE: If dECIMAL=3, the High Resolution Mode is not available. The display will show HI RESOLUTION and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it.
SERIAL PORT With SERIAL PORT displayed, press the PRINT/ENTER key. The display will change to RS232 MOdE. Proceed to the RS232 Port Output Mode parameter. Otherwise, press the ID/HEIGHT key to advance to the USB PORT menu. RS232 MOdE (Rs232 Port Output Mode) The display will show RS232 MOdE and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys, enter a new value and then press the PRINT/ENTER key to save it.
USb PORT With USb PORT displayed, press the PRINT/ENTER key. The display will change to USb MOdE. Proceed to the USB Port Output Mode parameter. Otherwise, press the ID/HEIGHT key to advance to the TIME SETTINGS menu. NOTE: If an Ethernet or WiFi option card is installed, pressing the ID/HEIGHT key will advance to the ETHERNET/WIFI menu. USb MOdE (USB Port Output Mode) The display will show USb MOdE and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it.
ETHERNET/WIFI – (If Option Card Installed) With ETHERNET/WIFI displayed, press the PRINT/ENTER key. The display will change to ETHERNET MOdE. Proceed to the Ethernet Port Output Mode parameter. Otherwise, press the ID/HEIGHT key to advance to the TIME SETTINGS menu. ETHERNET MOdE (Ethernet Port Output Mode) The display will show ETHERNET MOdE and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it.
TIME SETTINGS With TIME SETTINGS displayed, press the PRINT/ENTER key. The display will change to YEAR [YYYY]. Proceed to the YEAR parameter. Otherwise, press the ID/HEIGHT key to advance to the SCALE menu. YEAR [YYYY] (Year) The display will show YEAR [YYYY] and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys, enter a new value and then press the PRINT/ENTER key to save it. Allowable values are 2012 through 2099.
24 HOUR TIME (Use 24 Hour Time Format) The display will show 24 HOUR TIME and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise use the numeric keys 0/NO or 1/YES and enter a new setting. Press the PRINT/ENTER key to save it. Allowable values are YES or no. YES = Use 24 Hour Time Format (3PM displays 15:00) no = Use 12 Hour Time Format (3PM displays 3:00) Calibration Setup is Complete The Calibration Setup process has been completed.
42 8555-M512-O1 Rev A 03/13
SETUP REVIEW (Non-Protected Setup Parameters) The MedVue Model MV1 Medical Weight Analyzer allows several operational parameters to be reviewed and changed as necessary without having to remove the end cap and pressing the calibration switch to enter the Calibration Setup mode. To Enter Setup Review With the MV1 on and in the Gross weight mode: 1. Press and hold the CLEAR ( ) key for approximately 3 seconds. 2. The display will change to show SETUP REVIEW. 3. Release the CLEAR ( ) key. 4.
SERIAL PORT RS232 MOdE Set the mode of the RS232 serial port bAUd RATE Set baud rate for RS232 serial port (If RS232 MOdE = 1, 2 or 3) PRINTER TYPE Set the type of printer output (If RS232 MOdE = 3) End OF PRINT Set Number of End Of Print linefeeds (If RS232 MOdE = 3) USb PORT USb MOdE Set the mode of the USB port PRINTER TYPE Set the type of printer output (If USb MOdE = 3) End OF PRINT Set Number of End Of Print linefeeds (If USb MOdE = 3) ETHERNET/WIFI ETHERNET MOdE If option card is prese
ERROR AND STATUS DISPLAYS Display ERROR Meaning General error, invalid keypad entry was attempted. ERR OVERFLOW Attempting to display a negative number greater than –9,999 or a positive number greater than 99,999 ERR ZERO LIM Indicates an attempt to zero a weight outside scale zero range. Refer to the Calibration Setup ZERO Menu, ZERO LIMIT (Two Percent Zero Tracking Range Limit) parameter.
BEFORE YOU CALL FOR SERVICE Problem Display does not turn on Possible Solutions AC Operation: Is AC power supply fully inserted into wall receptacle? Check wall receptacle for proper AC power. Try another electrical appliance in same receptacle, does it work? Check circuit breaker. Has there been power failure? Battery operation: Check if batteries are installed and correctly.
SECURITY SEAL INSTALLATION If your metrology laws require the use of physical sealing, a lead and wire security seal can be installed on the MV1 to prevent the end cap from being removed from it to gain access to the calibration switch thereby preventing unauthorized access to the calibration adjustments. NOTE: In order to install the lead and wire security seal, the two screws securing the left end cap (as viewed from the MV1 front) MUST be replaced with two drilled screws.
PARTS IDENTIFICATION Item Qty Part Number Description 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 17 18 19 20 1 1 1 7 4 1 1 1 1 4 7 4 1 1 2 1 1 1 1 1 593GR986 6007-0009 6021-0423 6021-2069 6021-6008 6024-0039 6560-1125 6600-0659 6610-1233 6610-2000 6680-0004 6680-0052 8555-B509-08 8555-B510-08 8555-C213-08 8555-D501-0A 8555-D502-08 8555-D505-08 8555-D506-08 MV1PWR SERIAL TAG ASSY BOLT HEX HD SCW PAN-HEAD MACHINE-SCW 04-40X.250 SCW TRUSS HEAD SHEET METAL #6 X .50 SCW TRUSS HEAD SHEET-METAL #6 X.
8555-M512-O1 Rev A 03/13 49
APPENDIX A TOP PLATE MOUNTING CONFIGURATIONS “Type One” Uses Center Mounting Hole “Type Three” Uses Front Mounting Hole Center Mounting Hole Front Mounting Hole Models using “Type One” 6495 6437 6437DHR 6437KGDHR 6438 6439 6439KG 6439M 6447 6449 6449M 50 Models using “Type Three” 8430 8430M 8431 8431M 8437 8555-M512-O1 Rev A 03/13
“Type Four” “Type Five” Models using “Type Four” 6550 Models using “Type Five” 6854DHR 6857DHR 6857KGDHR “Type Desk/Wall Mount” Models using “Desk/Wall Mount” 6500 BRW1000 CD-1000D CR-500D FHD-133-II FHD-143-II FHD-144-II FHD-164-II GP-400-MV1 GP-600-MV1 GP-600-RP 4.7” 2.5” WALL MOUNTING Mount using at least two (2) #10 screws per dimensions on adjacent figure.
STATEMENT OF LIMITED WARRANTY Detecto Scale warrants its equipment to be free from defects in material and workmanship as follows: Detecto warrants to the original purchaser only that it will repair or replace any part of equipment which is defective in material or workmanship for a period of one (1) year from date of shipment. Detecto shall be the sole judge of what constitutes a defect.
STATEMENT OF LIMITED WARRANTY Conditions Which Void Limited Warranty This warranty shall not apply to equipment which: A.) Has been tampered with, defaced, mishandled or have had repairs and modifications not authorized by Detecto. B.) Has had serial number altered, defaced, or removed. C.) Has not been grounded according to Detecto’s recommended procedure. Freight Carrier Damage Claims for equipment damaged in transit must be referred to the freight carrier in accordance with freight carrier regulations.
Cardinal Scale Mfg. Co. PO Box 151, Webb City, MO 64870 Telephone: 417-673-4631 FAX: 417-673-5001 www.detectoscale.com Printed in USA 8555-M512-O1 Rev A 03/13 54 Technical Support: 866-254-8261 E-mail: techsupport@cardet.