ERROR AND STATUS DISPLAYS
Display Meaning
(1+
Motion is present when attempting to perform one of the following
operations: Power Up or Zero Weight Display.
A7A
Attempting to display a negative number less than -9,999 or a positive
number greater than 99,999.
$a%
Scale weight exceeds scale capacity.
".3
Indicates improper stored calibration data, calibration is necessary.
a."
The analog to digital circuit has failed. Consult scale serviceman.
"%a
The microprocessor has stopped receiving the signal from the analog
circuit. To correct, remove load from the scale platform and power the
indicator off and on.
<:0
Indicates an attempt to zero a weight outside the scale zero range. (See
Four Percent Zero Tracking Range Limit).
07a3<
This indicates the batteries are low. Replace or remove and recharge the
Ni-Cad batteries.
Refer to the 758CSV Series Operation Manual, 8555-M433-O1, for additional information.
CARE AND CLEANING
DO NOT subject scale platform to sudden shocks.
AVOID areas where scale might be exposed to moisture.
DO NOT submerge in water or spray water directly on scale platform or display indicator.
DO NOT use an abrasive cleaner, acetone or other volatile solvents for cleaning.
DO clean scale using a soft cloth dampened with a mild detergent. Display indicator may
be removed and washed with a soft damp cloth and a mild detergent.