Specifications
5.6.4. Test Procedure
a). The equipment was configured and connected to satisfy its functional requirements. It shall be
placed on the GRP with the interposition of a 0.1m-thick insulating support.
b). The equipment cabinets shall be connected to the safety earth directly on the GRP via the earth
terminal of the EUT.
c). The power supply, input and output circuits shall be connected to the sources of power supply,
control and signal.
d). The cables supplied or recommended by the equipment manufacturer shall be used. 1 meter of all
cables used shall be exposed to the magnetic field.
5.6.5. Test Result
Model Number I-7520A
Test Item Power Frequency Magnetic Field
Test Mode Mode 1
Date of Test 2014/02/07 Test Site TE08
Polarization Frequency (Hz)
Magnetic Strength
(A/m)
Performance Criterion Result
X Orientation 50 1 A B PASS
Y Orientation 50 1 A B PASS
Z Orientation 50 1 A B PASS
Note:
Criterion A : Operate as intended during and after the test
Criterion B : Operate as intended after the test
Criterion C : Loss/Error of function
Additional Information
EUT stopped operation and could / could not be reset by operator at dBuV (V) at frequency
MHz.
No false alarms or other malfunctions were observed during or after the test.
The acceptance criteria were met, and the EUT passed the test.