Specifications

5.5. Conducted Susceptibility (CS)
5.5.1. Test Specification
EN 61000-4-6
Environmental Phenomena Units Test Specification Performance Criterion
Signal Ports and Telecommunication Ports
MHz 0.15-80
V (rms, Un-modulated) 3
Radio-Frequency
Continuous Conducted
% AM (1kHz) 80
A
Input DC Power Ports
MHz 0.15-80
V (rms, Un-modulated) 3
Radio-Frequency
Continuous Conducted
% AM (1kHz) 80
A
Input AC Power Ports
MHz 0.15-80
V (rms, Un-modulated) 3
Radio-Frequency
Continuous Conducted
% AM (1kHz) 80
A
EUT tested in accordance with the specifications given by the standard of EN 61000-4-6.
Step : 1%
Step time : 3 Second
5.5.2. Test Instrument
Equipment Manufacturer Model Number Serial Number Cal. Date Remark
Signal Line Coupling
Decoupling Network
FCC FCC-801--T2-RJ11 8017 07/15/2013
(1)
Signal Line Coupling
Decoupling Network
FCC FCC-801--T4-RJ45 8018 07/15/2013
(1)
Signal Line Coupling
Decoupling Network
FCC
FCC-801-M2/M3-16A
8030
8030 07/15/2013
(1)
EM Injection Clamp FCC F-203I-23MM 8576 07/15/2013
(1)
NRP-Z91 POWER
SENSOR
R&S NRP-Z91 100613 07/15/2013
(1)
Amplifiers ar 75A250A 328729 N.C.R. -----
De-coupling Network FCC
F-203I-23MM-
DCN
8234 N.C.R. -----
Test Site ATL TE08 TE08 N.C.R. -----
Remark:
(1)
Calibration period 1 year.
(2)
Calibration period 2 years.
NOTE: N.C.R. = No Calibration Request.
A Test Lab Techno Corp.
Report Number: 1402CE13-01
61 of 153