Specifications
1.2. Measurement Uncertainty
Conducted Emission
The measurement uncertainty is evaluated as ± 2.02 dB.
Conducted Emissions (Telecommunication Ports)
The measurement uncertainty is evaluated as ± 2.02 dB.
Radiated Emission
The Vertical measurement uncertainty of 30MHz - 1GHz is evaluated as ± 3.62 dB.
The Horizontal measurement uncertainty of 30MHz - 1GHz is evaluated as ± 3.98 dB.
The Vertical measurement uncertainty of 1GHz - 6GHz is evaluated as ± 3.07 dB.
The Horizontal measurement uncertainty of 1GHz - 6GHz is evaluated as ± 3.11 dB.
Harmonic Current Emission
The measurement uncertainty is evaluated as ± 1.2 %.
Voltage Fluctuations and Flicker
The measurement uncertainty is evaluated as ± 1.5 %.
Electrostatic Discharge
As what is concluded in the document from Note2 of clause 5.4.6.2 of ISO/IEC 17025: 2005[E], the
requirements for measurement uncertainty in ESD testing are deemed to have been satisfied, and the
testing is reported in accordance with the relevant ESD standards. The immunity test signal from the ESD
system meet the required specifications in IEC 61000-4-2 through the calibration report with the calibrated
uncertainty for the waveform of voltage and timing as being 1.52 % and 2.69%.
Radiated susceptibility
As what is concluded in the document from Note2 of clause 5.4.6.2 of ISO/IEC 17025: 2005[E], the
requirements for measurement uncertainty in RS testing are deemed to have been satisfied, and the testing
is reported in accordance with the relevant RS standards. The immunity test signal from the RS system
meet the required specifications in IEC 61000-4-3 through the calibration for the uniform field strength and
monitoring for the test level with the uncertainty evaluation report for the electrical filed strength as being
2.65 dB.
Electrical fast transient/burst
As what is concluded in the document from Note2 of clause 5.4.6.2 of ISO/IEC 17025: 1999[2], the
requirements for measurement uncertainty in EFT/Burst testing are deemed to have been satisfied, and the
testing is reported in accordance with the relevant FT/Burst standards. The immunity test signal from the
FT/Burst system meet the required specifications in IEC 61000-4-4 through the calibration report with the
calibrated uncertainty for the waveform of voltage. Frequency and timing as being 1.57% and 2.73%.
A Test Lab Techno Corp.
Report Number: 1402CE13-01
6 of 153