Specifications

5.4. Surge
5.4.1. Test Specification
EN 61000-4-5
Item Environmental Phenomena Units Test Specification Performance Criterion
Signal Ports and Telecommunication Ports(See 1) and 2) )
Surges
Line to Ground
Tr/Th us
kV
1.2/50 (8/20)
± 1
B
Input DC Power Ports
Surges
Line to Ground
Tr/Th us
kV
1.2/50 (8/20)
± 0.5
B
Input AC Power Ports
Surges
Line to Line
Line to Ground
Tr/Th us
kV
kV
1.2/50 (8/20)
± 1
± 2
B
5.4.2. Test Instrument
Equipment Manufacturer Model Number Serial Number Cal. Date Remark
EMC Immunity Tester EMC-PARTNER AG TRANSIENT 2000IN6 952 02/06/2014
(1)
Test Site ATL TE08 TE08 N.C.R. -----
Remark:
(1)
Calibration period 1 year.
(2)
Calibration period 2 years.
NOTE: N.C.R. = No Calibration Request.
5.4.3. Setup
A Test Lab Techno Corp.
Report Number: 1402CE13-01
57 of 153