Specifications
5.2.5. Test Result
Model Number I-7520A
Test Item Radiated Susceptibility
Test Mode Mode 1
Date of Test 2014/02/10 Test Site TE07
Frequency
(MHz)
Polarity Azimuth
Field Strength
(V/m)
Performance
Criterion
Result
80 ~ 1000 H 0 3 A B PASS
80 ~ 1000 V 0 3 A B PASS
80 ~ 1000 H 90 3 A B PASS
80 ~ 1000 V 90 3 A B PASS
80 ~ 1000 H 180 3 A B PASS
80 ~ 1000 V 180 3 A B PASS
80 ~ 1000 H 270 3 A B PASS
80 ~ 1000 V 270 3 A B PASS
Note: The testing performed is from lowest level up to the highest level as required by standard, but
only highest level is shown on the report.
Criterion A: Operate as intended during and after the test
Criterion B: Operate as intended after the test
Criterion C: Loss/Error of function
Additional Information
There was no observable degradation in performance.
EUT stopped operation and could / could not be reset by operator at V/m
at frequency
MHz.
No false alarms or other malfunctions were observed during or after the test.
A Test Lab Techno Corp.
Report Number: 1402CE13-01
52 of 153