Specifications
5.2. Radiated Electromagnetic Field (RS)
5.2.1. Test Specification
EN 61000-4-3
Environmental Phenomena Units Test Specification Performance Criterion
Enclosure Port
Test Frequency Range MHz 80-1000
RF Electromagnetic Field V/m(Un-modulated, rms) 3
Amplitude Modulated % AM (1kHz) 80
A
EUT tested in accordance with the specifications given by the standard of EN 61000-4-3.
Step : 1%
Step time : 3 Second
5.2.2. Test Instrument
Equipment Manufacturer Model Number Serial Number Cal. Date Remark
SMB 100A SIGNAL
GENERATOR
R&S SMB100A 100724 03/06/2013
(2)
NRP-Z91 POWER
SENSOR
R&S NRP-Z91 100611 07/15/2013
(1)
NRP-Z91 POWER
SENSOR
R&S NRP-Z91 100612 07/15/2013
(1)
NRP POWER METER R&S NRP 101591 07/15/2013
(1)
Solid State Power
Amplifier
BONN ELEKTRONIK
BLWA
0830-160/100/40D
87050 N.C.R. -----
Signal Generator
Module
R&S SM300 Module 102209 N.C.R. -----
Broad-Band Horn
Antenna
Schwarzbeck
Mess-Elektronik
BBHA 9120 BBHA 9120 E388 N.C.R. -----
Test Site ATL TE07 888009 N.C.R. -----
Remark:
(1)
Calibration period 1 year.
(2)
Calibration period 2 years.
NOTE: N.C.R. = No Calibration Request.
A Test Lab Techno Corp.
Report Number: 1402CE13-01
50 of 153