Specifications
4 Emission Test
4.1. Conducted Emission Measurement
4.1.1. Limit
A.C. Mains Conducted Interference Limit:
Class A Equipment
(dBuV)
Class B Equipment
(dBuV)
Frequency
(MHz)
Quasi-peak Average Quasi-peak Average
0.15 - 0.5 79 66 66 - 56 56 - 46
0.50 - 5.0 73 60 56 46
5.0 - 30.0 73 60 60 50
Note: (1) The lower limit shall apply at the transition frequencies.
(2) The limit decreases in line with the logarithm of the frequency in the range 0.15 to 0.50 MHz.
Telecommunication Port Conducted Interference Limits:
Class A Equipment Class B Equipment
Voltage Limit
(dBμV)
Current Limit
(dBμA)
Voltage Limit
(dBμV)
Current Limit
(dBμA)
Requirement
(MHz)
QP Avg. QP Avg. QP Avg. QP Avg.
0.15 to 0.50 97 to 87 84 to 74 53 to 43 40 to 30 84 to 74 74 to 64 40 to 30 30 to 20
0.50 to 30 87 74 43 30 74 64 30 20
4.1.2. Test Instruments
Equipment Manufacturer Model Number Serial Number Cal. Date Remark
Test Receiver R&S ESCI 100367 06/06/2013
(1)
LISN R&S ENV216 101040 03/04/2013
(1)
LISN R&S ENV216 101041 03/04/2013
(1)
T-LISN FCC FCC-TLISN-T2-02 20574 04/19/2013
(1)
T-LISN FCC FCC-TLISN-T4-02 20529 04/19/2013
(1)
T-LISN TESQ ISN-T8 34413 04/25/2013
(1)
Test Site ATL TE02 TE02 N.C.R. -----
Remark:
(1)
Calibration period 1 year.
(2)
Calibration period 2 years.
NOTE: N.C.R. = No Calibration Request.
A Test Lab Techno Corp.
Report Number: 1402CE13-01
17 of 153