Product Manual
0064-0085-0M Rev B y 6880 Owner’s Manual 18
SCALE SETUP, CONT.
NOTE: The following two parameters are not used with the Model 6880 scale. Press
the
key to save the setting displayed, and then proceed to the next prompt.
;/ (Sonar Height Calibration)
The display will show ;! and the current setting 06. Press the
key to save the
current setting and proceed to the next prompt.
*`& (SENSOR HEIGHT)
The display will show *`; and the current setting. Press the
key to save the current
setting and proceed to the next prompt.
#! (HOLD TIME)
This setting is used by the scale to hold a stable patient weight for a desired amount of time.
For example, if a value of 5 seconds is used, then when the scale locks onto a stable
patient weight, it will remain locked for 5 seconds before automatically releasing the weight.
The display will show 6/3 and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to select a new value and then press
the
key to save it and proceed to the next prompt.
This setting controls the auto-locking feature. Allowable values for hold time are 0 to 20.