Specifications

8200-M411-O1 Rev D y 210/215 Installation & Technical 33
SETUP AND CALIBRATION, CONT.
If SMA is selected, the data will be transmitted in the following format:
<lf><s><r><n><m><f><xxxxxx.xxx><uuu><cr>
Where:
lf = Line Feed
s = Flags Z= center of Zero, O = Overcap, E = zero Error,
e = weight not currently being displayed
r = Range 1, 2, 3, ...
n = Mode G = Gross, T = Tare, N = Net
m = Motion M = Motion, " "(blank) = no motion
f = Custom Custom flag
xxxxxx.xxx =
Weight Six digits with decimal point
uuu = Units ton, lb , l/o, oz , t , kg , g
cr = Carriage Return (hex 0D)
If SB-400
*
or Computer is selected, the data will be transmitted in the following format:
<s><xxxxxx><d><uu><m><cc><cr>
Where:
s = Sign "-" = negative, " " (blank) = positive
xxxxxx.xxx =
Weight Six digits
d = Decimal point Added to string if enabled in setup
uu = Units tn, lb, l/o, oz, t, kg, g
m = Mode G = Gross, N = Net
cc = Weight Status OC = overcap
CZ = center of zero
MO = motion
ee = weight not currently being displayed
cr = Carriage Return (hex 0D)
*
The SB-80, SB-300 (multiple displays not supported) and WinVRS use the SB-400 format.
If SB-200 is selected, the data will be transmitted in the following format:
<cr><s><xxxxxx><d><c><uu><m> ETX
Where:
cr = Carriage Return (hex 0D)
s = Sign "-" = negative, " " (blank) = positive
xxxxxx.xxx =
Weight (with leading zeros)
d = Decimal point Embedded into weight (after weight dpp=0)
c = status m = motion
o = overcap
e = weight not currently being displayed
uu = Units tn, lb, l/o, oz, t, kg, g
m = Mode G = Gross, N = Net
ETX = End of TeXt (hex 03) MUST terminate ALL serial commands
If Rice Lake IQ355 is selected, the data will be transmitted in the following format:
<stx><s>xxxxxxx<u><m><s><cr><lf>
If AnDFV is selected, the data will be transmitted in the following format:
<hdr1>,xxxxx<uu><cr><lf>