Specifications

8200-M411-O1 Rev D y 210/215 Installation & Technical 10
INSTALLATION, CONT.
For completely isolated outputs, J11 must be open (positioned on one plug-in pin only or
removed) and the user must provide 12 to 24 VDC to the SRC pin and a ground return to the
load. The load must still be 200 ohms or greater.
To connect the control cable to the preset weight comparator/checkweigher logic level output
connector P10, first loosen the gland connector located on the right side on the back of the
210/215. Refer to Figure No. 8 for the exact location of the connector. Slip the cable through
the connector and into the enclosure. Remove 2 inches of the cable insulating jacket and then
1/4 inch of insulation from each of the internal wires (refer to Figure No. 4). Make the proper
terminations on terminal block P10. To terminate a wire, first press down on the terminal block
release bar, insert the wire into the terminal and remove pressure from the release bar locking
the wire in place.
Relay Board - (Optional)
The relay board (Cardinal p/n 8539-C062-0A) is mounted in the RB4-F external junction box
for use with the 210/215 Indicator. Connect devices to be controlled as shown in Figure No. 7.
The relays must
be configured to be on (closed) or off (open) at weights under the preset
weight then switch at the preset weight from on-to-off or off-to-on by setting the under weight
condition to on or off during setup and calibration or setup review. Refer to the Setup and
Calibration, “d out” (Digital Output) section of this manual for more information.
EXAMPLE: d oUt= 1,1
The PWC1 relay is on (closed) for weights under the preset weight and off (open) for weights
equal to or over the preset weight.
NOTE! All relays are the “normally-open” type that will open when power to
indicator is lost.
Figure No. 7