Specifications
8200-M242-O1 y 210-FE
Page 42
TROUBLESHOOTING, Cont.
Error Codes, Cont.
= a
UNITS key pressed in an attempt to perform a “unit” conversion that is not allowed.
CORRECTIVE ACTION: Determine the reason for the error display and take the
appropriate corrective action.
17<:a
NET key pressed with no stored tare weight value.
CORRECTIVE ACTION: Determine the reason for the error display and take the
appropriate corrective action.
$% (Over Capacity)
The load on the scale exceeds the scale capacity plus nine (9) divisions.
CORRECTIVE ACTION: Remove the over capacity load from the scale platform. May
indicate miscalibration.
A7A (Overflow)
The indicator is attempting to display a positive number greater than six (6) digits in length
or a negative number of more than five (5) digits.
CORRECTIVE ACTION: Return to Gross Weight mode and review Tare value. May
indicate miscalibration.
<773.a
UNITS key pressed in an attempt to perform a “unit” conversion where the interval would
have been greater than 50.
CORRECTIVE ACTION: Determine the reason for the error display and take the
appropriate corrective action.
(1+<3 (Unstable)
Motion is present when trying to power up, print, zero or perform a push button tare function.
CORRECTIVE ACTION: Wait for a stable weight display (STABLE annunciator on) before
performing any of these operations.