Specifications

8200-M242-O1 y 210-FE
Page 40
CALIBRATION “C” NUMBERS
The “C” numbers are displayed only during the Test mode operation by pressing the 4 key and
then the UNITS/TEST key. The "C" numbers are shown at the end of the test operation and
each number is displayed for approximately 4 seconds, allowing you to record them. Each
number may be up to three (3) digits in length. By recording these numbers you will be able to
return the indicator to its present calibration settings without using test weights simply by
entering the “C” numbers. Refer to the Setup and Calibration, """ section of this manual for
instructions on using the "C" numbers.
If any components have been changed that affect calibration and/or your scale
is used in a commercial application and must be "Legal for Trade" you can not
use "C" numbers to re-calibrate.
ACCUMULATORS
To view the NET accumulator:
1. Press the 4 key and then the NET/GROSS key.
2. Press the 4 key to return to normal operation.
To print the NET accumulator:
1. Press the 4 key, the NET/GROSS key and then the PRINT key
2. The indicator will return to normal operation when printing has been completed.
To clear (zero) the NET accumulator:
1. Press the 4 key, the NET/GROSS key and then the ZERO key
2. Press the 4 key to return to normal operation.
To view the Gross accumulator:
1. Press the 4 key and then the NET/GROSS key twice.
2. Press the 4 key to return to normal operation.
To print the Gross accumulator:
1. Press the 4 key, the NET/GROSS key twice and then the PRINT key
2. The indicator will return to normal operation when printing has been completed.
To clear (zero) the Gross accumulator:
1. Press the 4 key, the NET/GROSS key twice and then the ZERO key
2. Press the 4 key to return to normal operation.
To view the Count accumulator:
1. Press the 4 key and then the COUNT/SAMPLE key.
2. Press the 4 key to return to normal operation.
To print the Count accumulator:
1. Press the 4 key, the COUNT/SAMPLE key and then the PRINT key
2. The indicator will return to normal operation when printing has been completed.
To clear (zero) the Count accumulator:
1. Press the 4 key, the COUNT/SAMPLE key and then the ZERO key
2. Press the 4 key to return to normal operation.