Specifications
8200-M242-O1 y 210-FE
Page 32
SETUP AND CALIBRATION, Cont.
Single-Point for Span Only (Last Zero) Calibration
"? – FIRST CALIBRATION WEIGHT
The display will show "?. This is the first of two calibration steps. This weight is
the TEST WEIGHTS / TEST LOAD.
y Zero the scale and then place the weights on the scale platform.
y Using the numeric keys, input the value of the calibrated test weights / test load, then
press the ENTER key. NOTE! When entering values for "?, the digits start
displaying on the right side of the display and proceed to the left. When large values
are used (more than 3 digits), the "? prompt will automatically scroll off the left
side of the display to show the additional digits on the right as they are entered.
y Starting at the left and proceeding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
display will show: "?.
CAL2= - SECOND CALIBRATION WEIGHT
The display will show "?. This is the second of two calibration steps.
y Remove the weights on the scale platform and then press the ZERO key.
y The display will advance to +.7.
Single-Point for Zero Only (Only Zero) Calibration
"? – FIRST CALIBRATION WEIGHT
The display will show "?. This is the first of two calibration steps.
y Insure the scale is empty.
y Press the ENTER key.
y Starting at the left and proceeding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
display will show: "?.
"? - SECOND CALIBRATION WEIGHT
The display will show "?. This is the second of two calibration steps.
y Press the ZERO key.
y The display will advance to +.7.
Calibration “C” Numbers
1. With "? displayed, press the "diamond T” TARE key.
2. At the ? prompt, press the ENTER key to show the current value of the C1 number.
3. If the ? number displayed is acceptable, press the ENTER key again to save it.
4. Otherwise, use the numeric keys to enter a new ? number and then press the ENTER
key.
5. Repeat steps 2 through 4 for ?, ? and ?.
NOTE! If any components have been changed that affect calibration and/or
your scale is used in a commercial application and must be "Legal for Trade"
you cannot use the “C” numbers to re-calibrate.