Specifications

8200-M242-O1 y 210-FE
Page 31
SETUP AND CALIBRATION, Cont.
Dual-Point with Zero (First Zero) Calibration
"? – FIRST CALIBRATION WEIGHT
The display will show "?. This is the first of two calibration weights. This weight
could be ZERO (NO LOAD) or the TEST WEIGHTS / TEST LOAD.
y If the first calibration weight is to be ZERO (NO LOAD), press the ENTER key.
y If the first calibration weight is to be the TEST WEIGHTS / TEST LOAD, use the
numeric keys to input the value of the calibrated test weights. NOTE! When entering
values for "?, the digits start displaying on the right side of the display and
proceed to the left. When large values are used (more than 3 digits), the "?
prompt will automatically scroll off the left side of the display to show the additional
digits on the right as they are entered.
y Place the weights on the scale platform and then press the ENTER key.
y Starting at the left and proceeding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the display
will show: "?.
"? - SECOND CALIBRATION WEIGHT
The display will show "?. This is the second of two calibration weights. This
weight could be ZERO (NO LOAD) or the TEST WEIGHTS / TEST LOAD.
y If the second calibration weight is to be ZERO (NO LOAD), press the ENTER key.
y If the second calibration weight is to be the TEST WEIGHTS / TEST LOAD, use the
numeric keys to input the value of the calibrated test weights. NOTE! When entering
values for "?, the digits start displaying on the right side of the display and
proceed to the left. When large values are used (more than 3 digits), the "?
prompt will automatically scroll off the left side of the display to show the additional
digits on the right as they are entered.
y Place the weights on the scale platform and then press the ENTER key.
y Starting at the left and proceeding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily and then disappear, after which the
display will show: +.7.
Dual-Point without
Zero (False Zero) Calibration
"? – FIRST CALIBRATION WEIGHT
The display will show "?. This is the first of two calibration steps. This weight is
the TEST WEIGHTS / TEST LOAD.
y Place the weights on the scale platform.
y Using the numeric keys, input the value of the calibrated test weights / test load and
then press the ENTER key. NOTE! When entering values for "?, the digits start
displaying on the right side of the display and proceed to the left. When large values
are used (more than 3 digits), the "? prompt will automatically scroll off the left
side of the display to show the additional digits on the right as they are entered.
y Starting at the left and proceeding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
display will show: "?.
"? - SECOND CALIBRATION WEIGHT
The display will show "?. This is the second of two calibration steps.
y Remove the weights on the scale platform and then press the NET/GROSS key.
y Starting at the left and proceeding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the display
will show: +.7.