Specifications

8200-M242-O1 y 210-FE
Page 26
SETUP AND CALIBRATION, Cont.
%A3"? (Preset Print-On-Balance)
With only one Preset selected (4a$(<? 1 or 11), an additional prompt %A3"?(automatic
Print on Balance) will be displayed. If selected (%A3"?YES), when weight is above or equals
the preset value and all motion stops, weight will be printed (if a printer is attached). The
weight must
go below 50% of the preset value before another print operation can be
performed.
If the setting displayed is acceptable, press the ENTER key to save it. Otherwise, using the
numeric keys, 0/NO or 1/YES, enter the new setting and then press the ENTER key to save it.
%A3"? 1 (Yes) %A3"? 0 (No)
Automatic Print on Balance Enabled Automatic Print on Balance Disabled
%A3"? (Checkweigher Print-On-Accept)
With Checkweigher selected (4$(<? 4 or 14), an additional prompt %a3", Print on Balance,
(automatic print on accept) will be displayed. If selected (%A3"?YES), when the scale weight
is stable and in the accept range of the checkweigher, the weight will be printed (if a printer is
attached).
If the setting displayed is acceptable, press the ENTER key to save it. Otherwise, using the
numeric keys, 0/NO or 1/YES, enter the new setting and then press the ENTER key to save it.
%A3"? 1 (Yes) %A3"? 0 (No)
Automatic Print on Accept Enabled Automatic Print on Accept Disabled
+"%? (Sleep Mode Feature)
The Sleep Mode feature conserves battery power when the indicator remains unused for a
selected period of time. With the feature enabled, the load cell excitation will be reduced and
the display will be blank.
Press the ENTER key to show the current status of this feature. If a number other than 0 is
shown, this feature is selected and the number shown corresponds to the number of minutes
of a stable zero weight reading before the indicator enters the sleep mode. If the setting
displayed is acceptable, press the ENTER key again to save it. Otherwise, use the numeric
keys to enter a new value (0 to 10) and then press the ENTER key to store the new setting.
Note that entry of a 0 disables this feature.
a7? (Auto Shutoff)
The Automatic Shutoff feature will automatically turn the indicator off (when it is not in use)
after a predetermined period of inactivity to prolong battery life. To turn the instrument back on
you must press the ON / OFF key.
Press the ENTER key to show the current status for this feature. A number other than 0
indicates that the auto shutoff feature is enabled and the displayed number corresponds to the
number of minutes of stable weight displayed before the indicator is turned off automatically.
Note that a 0 indicates the feature has been turned off. If the setting displayed is acceptable,
press the ENTER key again to save it. Otherwise, use the numeric keys to enter a new value
(0 to 10) and then press the ENTER key to store the new setting.