Specifications

8200-M242-O1 y 210-FE
Page 23
SETUP AND CALIBRATION, Cont.
%++ (Password Y/N)
The %++ (Password Y/N) prompt determines whether a password is required to enter Setup
and Calibration on indicators programmed for international use. Note that this prompt is only
displayed when (+a?a (International) is selected.
With the display showing PASS, press the ENTER key to show the current value. If the setting
displayed is acceptable, press the ENTER key again to save it. Otherwise, using the numeric
keys, 0/NO or 1/YES, enter the new setting, then press the ENTER key to save it.
%++ (NO) %++ (YES)
Password protection is NOT needed.
Setup advances to the LFT= prompt.
Password protection is desired. The
next prompt will be %++?
Press the ENTER key to see the current password value. If the password displayed is
acceptable, press the ENTER key again to save it. Otherwise, using the numeric keys enter
the new password (up to 6 digits) and then press the ENTER key to save it.
"<? (Legal For Trade)
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, using the numeric keys, 0/NO or 1/YES, enter the
new setting and then press the ENTER key to save it.
"<?a "<?a
Interval Settings (*1<?) allowed
are: 1, 2, 5, 10 , 20, 50
Interval Setting (*1<?) is
selectable from 1 to 99.
NOTE! When both "<? and (+?, the followings results occur:
Scale must have between 100 and 10,000 divisions
':2? .5 or 0 to 3
Inhibit serial data during input
Disables COUNT key
2<5? mm/dd/yy
':0? no
29 + 4% to OC
NOTE! When "<? and (+?, the followings results occur:
(1;? 1
2<5? dd/mm/yy
':a? yes
29 + 9 grads to OC
PT printed with tare
Lamp test on power up
Password Operation
With the %++ prompt enabled (set to YES), anytime the operator tries to enter +5<=9, the
display will show the %++? prompt requiring the operator to enter the correct password. If
the wrong password is entered the indicator displays &&$& momentarily and returns to the
weight mode. Note that the password is not shown on the display when the operator is
entering it.
NOTE! It is recommended to write the password down and store it in a secure
location. If
the password is forgotten or lost and a change to the indicator setup is required, the
indicator must be reprogrammed. WARNING! Reprogramming the indicator will erase all
the contents of the Nov-Ram and memory.