Specifications

8200-M242-O1 y 210-FE
Page 21
SETUP AND CALIBRATION
Your Model 210-FE indicator has been thoroughly tested and calibrated before being shipped
to you. If you received the indicator attached to a scale, calibration is not necessary. If the
indicator is being connected to a scale for the first time or recalibration is necessary for other
reasons, proceed as indicated.
The calibration switch is located on a bracket on the inside of the enclosure rear panel. You
may gain access to this switch simply by removing the calibration switch access screw on the
rear panel. Refer to Figures No. 11 and 12.
Enter Setup Mode
With the 210-FE indicator ON;
1. Insert a small screwdriver or other tool
through the calibration switch access hole
on the rear panel. Refer to Figure No. 13.
2. Press and release the calibration switch.
3. The menu Setup will be displayed.
4. Continue to press and release the switch to
rotate through the beginning point for
entering the setup mode.
+<(% Setup Mode (starts at (+ prompt)
A4 Analog to Digital Filtering (starts at 4"<? prompt)
"a Calibration (starts at " prompt)
+*$ Serial Input/Output (starts at ( prompt)
%:.1< Print Tab Settings (starts at %$&' prompt)
+%1 Fine Span Adjustment
.a:+ Display high resolution weight mode
"77(< Key lock out function