Specifications

8200-M242-O1 y 210-FE
Page 8
INSTALLATION, Cont.
8. To terminate a wire, use a small flat blade
screwdriver and press down on the release
bar for the terminal. Insert the wire into the
terminal opening. Remove the screwdriver,
allowing the release bar to return to its
original position, locking the wire in place.
See Figure No. 5.
9. Repeat the procedure until all of the wires
are in place.
10. After all terminations have been made,
remove the excess cable from the enclosure.
11. Referring to Figure No. 6, fold the shield wire back over the plastic insert and then insert
the plastic insert (with the shield wire) into the gland connector.
12. The shield wire is secured when tightening the gland connector nut. See Figure No. 7.
13. Do not over-tighten the connector but make certain it is snug.
Load Cell Cable Connection (Standard Gland Connector)
The following instructions describe the load cell connection should it be desired to route the
load cell cable through a standard gland connector.
1. Remove the 16 acorn nuts securing the rear panel to the main housing.
2. Referring to Figure No. 2, choose a gland connector for the cable and then loosen it.
3. Slip the single cable from the load cell or load cell junction box through the gland connector
and into the enclosure.
4. Referring to Figure No. 4, remove 3 inches (not 5 to 6 inches) of the outer insulation jacket
5. Next, remove 1/4 inch of insulation from each of the 4 wires and shield (without sense
leads) or 6 wires and shield (with sense leads).
6. Referring to the table on the previous page (or on the circuit board) for terminal
connections, connect each of the wires to terminal block P7. Refer to Figure No. 9 for
terminal block location.
7. To terminate a wire, use a small flat blade screwdriver and press down on the release bar
for the terminal. Insert the wire into the terminal opening. Remove the screwdriver,
allowing the release bar to return to its original position, locking the wire in place.
8. Repeat procedure until all of wires are in place.
9. If a standard gland connector is used, the shield wire should be connected to terminal 4 of
the P1 terminal block connector.
Figure No.5
Figure No.7
Figure No.6