Specifications
8200-M024-O1 Rev H y 205 Installation & Technical
37
ERROR CODES, CONT.
99 (Error 1)
A program checksum mismatch has been detected.
CORRECTIVE ACTION
: Consult your scale service provider.
99 (Error 3)
Internal RAM failure.
CORRECTIVE ACTION
: Consult your scale service provider.
< (HuH?)
UNITS key pressed in an attempt to perform a unit conversion that is not allowed.
CORRECTIVE ACTION
: Determine the reason for the error display and take the
appropriate corrective action.
06;9`(no tarE)
NET key pressed with no stored tare weight value.
CORRECTIVE ACTION
: Determine the reason for the error display and take the
appropriate corrective action.
6$ (over Capacity)
The load on the scale exceeds the scale capacity plus nine (9) divisions.
CORRECTIVE ACTION
: Remove the over capacity load from the scale platform. May
indicate miscalibration.
@6@ (overflow)
The indicator is attempting to display a positive number greater than six (6) digits in length
or a negative number of more than five (5) digits.
CORRECTIVE ACTION
: Return to Gross Weight mode and review Tare value. May
indicate miscalibration.
;662- (too big)`
UNITS key pressed in an attempt to perform a unit conversion where the interval would
have been greater than 50.
CORRECTIVE ACTION
: Determine the reason for the error display and take the
appropriate corrective action.
'0*;2 (Unstable)
Motion is present when trying to perform a print function.
CORRECTIVE ACTION
: Wait for a stable weight display (STABLE annunciator on) before
performing any of this operation.