Specifications
8200-M024-O1 Rev H y 205 Installation & Technical
32
SETUP AND CALIBRATION, CONT.
`*$0 (*$0) - Fine Span Adjustment
NOTE! The *$0 and -`9* modes require a load of 10% of Capacity be on
the scale before adjustments can be made.
With the `*$0 prompt displayed after pressing the TARE/
ENTER key at the last Print
prompt:
With *$0 displayed, press the TARE/ENTER key. The display will change to show the
current setting 06. To skip the Fine Span Adjustment and return to the *;'$ menu, press
the TARE/
ENTER key again. To perform the Fine Span Adjustment, place a calibrated test
weight on the scale and press the ASTERISK/
UP ARROW key (display will change to (*)
and then the TARE/
ENTER key.
After pressing the TARE/
ENTER key, the display will change to show the amount of the test
weight and the annunciators will alternately flash off and on i.e. (all ON, weighing unit off,
then all OFF, weighing unit ON). Press the ASTERISK/
UP ARROW key to increase the span
OR press the UNITS/LEFT ARROW key to decrease the span. Press the ASTERISK key
to return to the previous prompt or press the TARE/
ENTER key to exit *$0 and return
to the *;'$ menu.
With the `*$0` prompt displayed after pressing the Calibration switch:
With `*$0 displayed, place a calibrated test weight on the scale and press the
TARE/
ENTER key. After pressing the TARE/ENTER key, the display will change to show the
amount of the test weight and the annunciators will alternately flash off and on (all ON,
weighing unit off, then all OFF, weighing unit ON). Press the ASTERISK/
UP ARROW key to
increase the span OR press the UNITS/LEFT ARROW key to decrease the span. Press
the ASTERISK/
UP ARROW key to return to the previous prompt or press the TARE/ENTER
key to exit `*$0 and return to the *;'$ menu.
-`9* - Display High Resolution Weight
With -`9* on the display, pressing the TARE/ENTER key will show the active weight in
"high resolution" mode (in 1/10 interval). Press the PRINT key to print the weight (followed by
the text TEST) via the selected printer output port enabled during setup and calibration. Press
the TARE/
ENTER key to return to the -`9`*`prompt. To exit the -`9`* mode, press
the calibration switch or cycle power (press the ON/OFF key twice).
!66'; - Key Lock Out Function
With !66'; on the display, pressing any key will display !63 (locked) or '0!63
(unlocked) for the current key state. Pressing a locked key during normal operation will results
in a 1/2 second display !63 and the key will be ignored. To exit the !66'; function, press
the calibration switch or cycle power (press the ON/OFF key twice).
NOTE! The menu selections -`9* and !66'; can only be selected using
the calibration switch.