Specifications

8200-M024-O1 Rev H y 205 Installation & Technical
28
SETUP AND CALIBRATION, CONT.
If WI110 is selected, the data will be transmitted in the following format:
<m><sp><s>xxxxx<sp><uu><cr><lf>
If number only is selected, the data will be transmitted in the following format:
xxxxxx<cr><lf>
If Toledo Short is selected, the data will be transmitted in the following format:
<stx><swa><swb><swc>xxxxxx<cr><sum>
Weight On Demand
If continuous output has not been selected for Serial Port 1 (Cont1=NO), the 205 indicator will
respond to a weight request (ENQ).
The host device (computer) sends:
ENQ - (hex 05)
The 205 will respond:
<s><xxxxxx><d><uu><m><cc><cr>
Where:
s = Sign "-" = negative, " " (blank) = positive
xxxxxx.xxx =
Weight Six digits
d = Decimal point Added to string if enabled in setup
uu = Units tn, lb, l/o, oz, t, kg, g
m = Mode G = Gross, N = Net
cc = Weight Status OC = overcap
CZ = center of zero
MO = motion
ee = weight not currently being displayed
cr = Carriage Return (hex 0D)
NOTE! The Weight On Demand function is not available for Serial Port 2.
60;> (Continuous Output Serial Port 2)
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, using the arrow keys, select a new setting then
press the ENTER key to save it.
60;>`(*` 60;>`06`
Continuous Output No Continuous Output
If 60;>`(* (Continuous Output) is selected, an additional prompt, ;($> will be
displayed.
;=$> (Continuous Output Format)
Press the TARE/
ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/
ENTER key again to save it. Otherwise, use the ASTERISK/UP ARROW key to
toggle to a new setting and then press the TARE/
ENTER key to save it. Allowable values are:
0 = SMA 3 = Rice Lake IQ355 6 = Number
1 = SB-400 4 = AnDFV 7 = Toledo Short
2 = SB-200 5 = WI110 8 = SB500 with Traffic Light
NOTE! See Continuous Output Serial Port 1, ;($> for description of output formats.