Specifications

8200-M024-O1 Rev H y 205 Installation & Technical
25
SETUP AND CALIBRATION, CONT.
Single-Point for Zero Only (Only Zero) Calibration
IMPORTANT! The functions of numeric keys are replaced by using the
UNITS/
LEFT ARROW and the ASTERISK/UP ARROW keys. The cursor location is
identified by the blinking character and can be advanced to the left to the next
position by pressing the UNITS/
LEFT ARROW key. Pressing the ASTERISK/UP
ARROW
key will change the blinking character to the next value.
!> – First Calibration Weight
The display will show !>. This is the first of two calibration steps.
Insure the scale is empty.
Press the TARE/
ENTER key.
Starting at the left and proceeding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
display will show: !>.
!> - Second Calibration Weight
The display will show !>. This is the second of two calibration steps.
Press the ZERO key.
The display will advance to *-6.
Calibration “” Numbers
1. With !> displayed, press the UNIT key.
2. At the > prompt, press the TARE/
ENTER key to show the current value of the  number.
3. If the > number displayed is acceptable, press the TARE/
ENTER key again to save it.
4. Otherwise, use the ASTERISK/
UP ARROW and UNITS/LEFT ARROW keys to enter a new
> number, then press the TARE/
ENTER key.
5. Repeat steps 2 through 4 for >, > and >.
NOTE! If any components have been changed that affect calibration and/or
your scale is used in a commercial application and must be "Legal for Trade"
you cannot use the “” numbers to re-calibrate.