SETUP AND CALIBRATION, CONT.
`@`3 (`@`3) - Analog to Digital Filtering
3!;> (Digital Filtering)
With `@`3`(`@`3) displayed, press the TARE/
ENTER key. The display will change to
3!;>. Press the TARE/
ENTER key to show the current value. If the setting displayed is
acceptable, press the TARE/
ENTER key to save it. Otherwise, use the ASTERISK/UP ARROW
key to toggle to a new setting and then press the TARE/
ENTER key to save it. Allowable
values are: 0, 1, 2 or 3. Note, that if you select 3 (Custom Filtering) two additional prompts will
be displayed.
3!;>`
0 Disabled – NO Filtering
1 MINIMAL FILTERING (sample rate = 2)
2 MODERATE FILTERING (sample rate = 1)
3 CUSTOM FILTERING
NOTE! The prompts, > (Filter Level) and 2> (Break Range) will only
be displayed if you
selected 3 (Custom Filtering) for the 3!;> (Digital Filtering) prompt.
>`(Filter Level)
Press the TARE/
ENTER key to show the current setting for the filter level. The filter level is
a number from 1 to 99 that corresponds to the level of filtering with 99 being the greatest
filtering and 1 the least. To accept the value displayed, press the TARE/
ENTER key,
otherwise, use the ASTERISK/
UP ARROW key to toggle to a new setting and then press the
TARE/
ENTER key to save it.
2> (Break Range)
Press the TARE/
ENTER key to show the current setting for the break range. The break
range is a number from 1 to 255 that corresponds to the number of division change to
break out of the filtering. Press the TARE/
ENTER key to keep the displayed value or use
the ASTERISK/
UP ARROW key to toggle to a new setting and then press the TARE/ENTER
key to save it. Note that selecting 0 disables this feature.
*9> (Sample Rate)
Press the TARE/
ENTER key to show the current setting for the sample rate. The value
displayed is the sample rate in samples per second. Press the TARE/
ENTER key to save
the displayed value or use the ASTERISK/
UP ARROW key to toggle to a new value (1 to
100) and then press the TARE/
ENTER key to save it.
'0*> (Motion Range)
Press the TARE/
ENTER key to view the current setting for the range of motion detection. If the
displayed value is acceptable, press the TARE/
ENTER key to save it. Otherwise, use the
ASTERISK/
UP ARROW key to toggle to a new range (the number of divisions of change
permitted before indicating unstable) and then press the TARE/
ENTER key to save the new
setting. Allowable range values are: 0 through 99 divisions.
*> (Stable Count)
Press the TARE/
ENTER key to view the current setting for the number of consecutive stable
weight readings before indicating stable weight. This helps filter weight readings for stability
when trying to capture stable weight. If the displayed value is acceptable, press the
TARE/
ENTER key to save it. Otherwise, use the ASTERISK/UP ARROW key to toggle to a new
value and press the TARE/
ENTER key to save the new setting. Allowable values for the stable
count are: 3 through 255.