Specifications

SETUP AND CALIBRATION, CONT.
Hold Key Enable
The display will next indicate
HLd=
which is the prompt for the selection to enable or disable
the Hold (ASTERISK) key. This mode of operation will lock the 204 display and is used only in
noncommercial applications. Press the ASTERISK key to show the current setting, then press
the UNITS key to select 1 (1=YES, enable) or 0 (0=NO, disable). Press the PRINT key to save
the setting.
NOTE! This feature must be set to 0 (0=NO) for “Legal for Trade” applications.
Battery Type
With the display showing
niCAd=
press the ASTERISK key to show current setting and then
press UNITS to select the type of batteries to be used. NOTE! This setting may be revised
without having to enter the calibration mode.
niCAd=
0 Alkaline batteries - battery charging is DISABLED
1 NiCad or NiMH batteries - battery charging is ENABLED
2 Battery charging is ENABLED and FORCED ON. This selection
forces battery charging for NiCad or NiMH batteries that are
discharged. NOTE! After 15 hours of charging, the indicator will
automatically change the
niCAd=
setting back to a 1.
Once the correct value is shown press the PRINT key to save the setting.
CAUTION: Selecting 1 or 2, enables battery charging. DO NOT select 1 or 2
when using Alkaline batteries.
Power Up Zero
With the display showing
PUO=
press the ASTERISK key to show current setting, then press
UNITS to select 1 (1=YES, enable) or 0 (0=NO, disable). Press the PRINT key to save the
setting. NOTE! This setting may be revised without having to enter the calibration mode.
x If 1=YES is selected, the weight display will be reset to zero automatically on power up.
x If 0=NO is selected, the weight display will not be reset to zero.
Automatic Shutoff
The display will now show
ASH=
. This Automatic Shutoff feature will automatically turn the
204 off after a predetermined period of inactivity to prolong battery life. To turn the indicator
back on you must press the ON / OFF key.
Press the ASTERISK key to show the current setting. Use the UNITS key to select the
number (1 through 9) of minutes (time approximate) of inactivity before turning the 204 off. A 0
disables the Automatic Shutoff feature. Press the PRINT key to save the setting. NOTE! This
setting may be revised without having to enter the calibration mode.
8555-M313-O1 Rev G x 204 Installation & Technical
18