Specifications
SETUP AND CALIBRATION, CONT.
USA (Domestic or International)
The display will next indicate
USA=
. This is the prompt to select whether the 204 is used in the
USA (domestic) or outside the US (international). Press the ASTERISK key to show current
setting, then press the UNITS key to select 1 (1=YES, Domestic) or 0 (0=NO, International).
USA = 1 (Domestic) USA = 0 (International)
No Zero Limit +/- 2% Zero Limit
Tare Resolution = 1d (1 division) Tare Resolution = .25d (1/4 division)
Lamp test on power up enabled (display
segments 1 second on, 1 second off)
Once the correct value is shown press the PRINT key to save the setting.
Motion (Unstable) Range
The display will now show
Un5=
. Press the ASTERISK key to show the current setting.
Changes in weight exceeding the selected number of divisions will cause the STABLE weight
annunciator to turn off. Values from 1 to 9 divisions may be selected by pressing the UNITS
key. Once the correct value is shown press the PRINT key to save the setting.
Digital Filter Level Selection
The display will now show
FLt=
which is the prompt for the selection of the digital filtering
level. Your 204 will arrive with the factory filter setting (1=minimal) already entered. Please
check with your scale service technician should you wish to change the programmed filter level
and break range. Four levels of filtering are available. They are as follows:
0 = NO FILTERING 2 = MODERATE FILTERING
1 = MINIMAL FILTERING 3 = CUSTOM FILTERING
NOTE: Selection 3, Custom Filtering is used when 0, 1 or 2 are inadequate.
Press the ASTERISK key to show the current setting. Then press the UNITS key to select the
desired level of filtering. Press the PRINT key to save the setting.
F=
- Filter Level
If you select Custom Filtering, the 204 will display
F=
. Press the ASTERISK key to show the
current setting for the Filter Level. The filter level is a number from 1 to 16 that corresponds
to the level of filtering with 16 being the greatest filtering and 1 the least. Use the UNITS and
ASTERISK keys to select the filter level and then press the PRINT key to save the setting.
br=
- Break Range
Next, the 204 will display
br=
. Press the ASTERISK key to show the current setting for the
Break Range. The break range is a number from 1 to 64 that corresponds to the number of
division change to break out of filtering. Use the UNITS and ASTERISK keys to select the
break range value and then press the PRINT key to save the setting.
Sample Rate
The display will now show
Sr=
. Press the ASTERISK key to show the current setting. The
sample rate may be set from a minimum of 1 sample per second to a maximum of 10 samples
per second in one sample per second intervals. Press the UNITS key until the desired sample
rate is displayed. Press the PRINT key to save the setting.
8555-M313-O1 Rev G x 204 Installation & Technical