Specifications
changes in the “Performance Standard" column.
+19.00 and +21.80.
In step 2b, change “+14.2” to: between
In step 4b, change
+13.40 and +15.00.
+9.90 and +1120.
In step 2c, change “+17.6” to: between
In step 4c, change
+17.30 and +19.50.
+12.70 and +14.20.
“+10.50” to: between
“+13.00” to: between
In step 3b, change “+20.7” to: between
Add step No. 5 after step No. 4.
Equipment under
Test
equipment
test control
Test
Performance
Step No.
control settings
settings
procedure
standard
5
Sane as step No. 4
Same as step No. 1
a.
Same as step
a RED indicator
except: Selector
No. 4a.
lights.
switch: 19
b.
Same as step
b.
TS-443/U indi-
No.
4b.
cdes
between-
+13.00 and
+14.70 volts.
Page 43, paragraph 48c. Make the following
changes:
Add the following after the heading of sub-
paragraph c: The light indication change to
GREEN may occur instantaneously, but if it
does not the change should occur within 10 sec-
nds after required voltage limits are applied to
the test probe.
Make the following changes in the “Per-
fomance
Standard" column:
In step lb, change “2.85” to: between 2.60
and 3.30.
In step 2b, change “U3” to: between 1250
and 14.90.
In step 3b, change “0.49” to: between .450
and .520.
In step 4b, change ‘0.90” to: between .800
and 1.00.
In step 5b, change “0.225” to: between .200
and .250.
In step 6b, change “6.40” to: between 6.00
and 7.30.
Page 47, paragraph 49c. Make the following
changes in the “Performance standard” column:
In step No. lc, change “0.180” to: between
.125 and .250.
In step No. 2c, change “0.220” to: between
.200 and .250.
In step No. 3c, change “0.480” to: between
.170 and 210.
In step No. 4c, change “0.720” to: between
.650 and .810.
Page 51, paragraph 50c. Make the following
changes in the “Performance standard” column:
In step No. lb, change “0.27” to: between
0230 and .330.
In step No. 2b, change “0.130” to: between
.100 and .140.
In step No. 3b, change “0.225” to: between
.180 and .250.
In step No. 4b, change “0.014” to: between
.012 and .016.
Page 53, paragraph 51c. In “Performance stand-
ard” column, change subparagraph b to read:
411A indicates between .100 and .135 volt.
Page 54, paragraph 52, chart. Delete the chart
and substitute:
7