User Manual

3.17 SOFTWARE FLOW CHART
Detcon Model TP-624C Hydrogen Sulfide Sensor PG.25
P
GM1 (3) PGM2 (3)
NORMAL
OPERATION
P
GM1 (M)
P
GM2 (M)
PGM1 (3)
SET ALM1 @ ##PPM
I
NC
D
EC
PGM1 (3) PGM2 (M)
LEGEND
PGM1 - program switch location #1
PGM2 - program switch location #2
(M) - momentary pass of magnet
(3) - 3 second hold of magnet
(30) - 30 second hold of magnet
INC - increase
DEC - decrease
# - numeric value
A
UTO SPAN
S
ET ALARM 1 LEVEL
GAS RANGE V#.#
PGM1 (3) PGM2 (M)
V
IEW PROG STATUS
A
LM1 SET @ ##PPM
ALM2 SET @ ##PPM
485 ID SET @ #
P
GM1 (M)
P
GM2 (M)
PGM1 (3)
CalLevel @ ##PPM
I
NC
D
EC
PGM1 (3) PGM2 (M)
S
ET CAL LEVEL
P
GM2 (3)
P
GM2 (3)
P
GM2 (3)
SENSOR LIFE ##%
A
LM1 (Firing Direction)
ALM1 (Latch State)
ALM1 (Energize State)
ALM2 (Firing Direction)
ALM2 (Latch State)
ALM2 (Energize State)
FLT (Latch State)
FLT (Energize State)
PGM1 (3) PGM1 (M)
P
GM2 (30) PGM2 (M)
ALARM RESET
A
LARM RESET
P
GM1 (M)
P
GM2 (M)
PGM1 (3)
SET ALM2 @ ##PPM
I
NC
D
EC
PGM1 (3) PGM2 (M)
S
ET ALARM 2 LEVEL
P
GM2 (3)
CAL LEVEL @ ##PPM
RANGE @ 0-###PPM
HEATER @ #.##VDC
PGM1 (3) PGM2 (M)
PGM2 (3)
APPLY 10 PPM H2S
LINEARIZE SENSOR
PGM1 (M)
PGM2 (M)
PGM1 (3)
INC
DEC
PGM1 (3) PGM2 (M)
PGM2 (3)
HEATER @ ##VDC
SET HEATER VOLTS
APPLY 50 PPM H2S
PGM1 (M) PGM2 (M)
PGM1 (3)
INC
DEC
PGM1 (3) PGM2 (M)
PGM2 (3)
RANGE @ 0-###PPM
SET RANGE
Shipping Address: 3200 A-1 Research Forest Dr., The Woodlands, Texas 7381
Mailing Address: P.O. Box 8067, The Woodlands, Texas 77387-8067
phone 888-367-4286, 713-559-9200 • fax 281-292-2860 • www.detcon.com • sales@detcon.com