Owner manual
3
Temperature
Model
RFD 1-42-2E
Time to Temperature 25ºC – 100ºC
(approximate minutes with 25ºC – 200ºC
no load) 25ºC – 343ºC
4
10
35
Recovery Time Door Open 1 Min. 100ºC
(approximate minutes with 200ºC
no load) 343ºC
<1
2
4
Temperature Uniformity at 100ºC*
200ºC*
343ºC*
±1ºC
±2ºC
±4ºC
Minimum Operating Temperature dampers open
(Approximate w/20ºC ambient) dampers closed
40ºC
50ºC
Control Stability
±0.3ºC
Repeatability
±0.5ºC
Cooldown Time 343 – 50ºC dampers open
(approximate 343 – 75ºC dampers closed
minutes with no load)
46
110
* Figures are based on actual tests in an empty oven. Uniformity can vary slightly