Manual
6
Temperature
RBC1-50
Air Atmosphere
RBC1-50
Nitrogen Atmosphere
Time to Temperature 25°C - 80°C
(approximate minutes with 25°C - 125°C
no live load) (1) 25°C - 150°C
25°C - 260°C
4
10
14
34
4
10
14
30
Recovery Time to 125°C with door open 1 Min.
>1 Min. >1 Min.
Temperature Uniformity at (2) 125°C*
260°C*
<±1°C
±1°C
<±1°C
±1°C
Operating Temperature Range
(Based on 25°C ambient, and dampers fully open)
50° - 260°C 15° - 260°C
Control Stability (3)
±0.5°C per 5°C
∆
Repeatability ±0.5°C
Cooldown Time 125° - 60 °C
40 Min. 11 Min.
Dissipation/Mean Temperature Rise (4) 80°C
125°C
150°C
800 Watts/4°C
1,300 Watts/6°C
1,400 Watts/6°C
(1) Time to temperature will be slightly longer for lower voltages due to heater kW derating (empty oven).
(2) Figures are based on a nine-point survey in an empty oven. Uniformity can vary slightly depending
on unit and operating conditions.
(3)
∆
In control stability figures represent the change in ambient temperature.
(4) Dissipation is reduced by 17% on 50-Hertz operation. Temperature gradient/mean temperature rise is
with an evenly distributed live load.
Note: Testing performed on 240V-1PH-60HZ.