User Manual

25
NOTE: ### or ##.# represents a numeric value or parameter.
Tune Mode Parameter Calibration
Press Display Factory Default
Setting
Actual
Setting
Press Adjustable Range
key P-1 ### 5 + or - keys 0E C to 500E C
32E F to 932E F
key I-1 ### 30 + or - keys 0 to 999
seconds/repeat
key D-1 ### 0 + or - keys 0 to 999
degrees/second
key DEG - C C + key for C or - key for F.
key SPL - ### maximum designed
operating
temperature¹
+ or - keys 0E C to 500E C
32E F to 932E F
key CZO - ##.# can vary + or - keys to change
parameter to 0.0
-99.9 to 99.9
degrees
key DIS
HHMM
HHMM + key for HHMM (hours
and minutes) or the - key
for MMSS (minutes and
seconds).
key BEEP ON ON + key for ON or the - key
for OFF.
¹ 204E C/400E F, 260E C/500E F, 343E C/650E F
e. Press the Manual key. The display reads [MANUAL ].
f. Allow the controller a thirty (30) minute warm up time before proceeding to
the step #13 - CAL MODE.
12. Press the Cal key. The display reads [CAL--MODE].
13. Press the key. The display reads [CODE ***].
14. Press the following key sequence: -, +, +, -, +, -.
The display reads [HCAL 250].