5.6.2. Working with Calibration Offsets
Calibration offsets allow configuration for oven temperature variations. Enter calibration offsets
through the Configuration Menu and Input Configuration.
5.6.2.1. Access the Configuration Menu
To access the Configuration Menu and all sub-menus, the controller must be in its Cycle
Complete or Stopped state (that is, not running). If the controller is not in the Stopped state, press
to stop the controller before proceeding.
1. Press and together from the Select a Mode screen to access the Main Menu.
2. From the Main Menu, navigate to and select (highlight) the Configuration Menu. Press
.
3. A prompt will appear to enter an unlock code. Enter the correct code (the default code is
0010) with and then press to access the Configuration Menu.
5.6.2.2. Working with Input Configuration
Input Configuration allows you to configure items associated with the Control Input Setup,
Control Input Cal., High Limit Input Setup and High Limit Input Cal. Control calibration offset
parameters are located in the Control Input Cal.
To work with Input Configuration:
1. Navigate from the Configuration Menu to and highlight Input Configuration. Press
.
2. Navigate to and select Control Input Cal. Press .
a. Three options present: Factory Calibration, Single Point Calibration, Two Point
Calibration.
b. Only one option can be active at any given time. The highlighted choice becomes the
active choice.
3. Navigate to and select the desired choice. Press to configure the settings within
that choice.