39
OUTPUT SPECIFICATIONS
OUTPUT 1 (Limit Relay)
Relay
Contact Type: SPDT
Rating: 5A resistive at 120/240V AC
Lifetime: > 100,000 operations at rated voltage/current
Isolation: Inherent
OUTPUT 2 (Alarm or Annunciator)
Relay
Contact Type: SPDT
Rating: 2A resistive at 120/240V AC
Lifetime: > 500,000 operations at rated voltage/current
Isolation: Inherent
OUTPUT 3 (Alarm, Annunciator, or Retransmit)
General
Types Available: Relay or DC linear (retransmit only)
Relay
Contact Type: SPDT
Rating: 2A resistive at 120/240V AC
Lifetime: > 500,000 operations at rated voltage/current
Isolation: Inherent
DC
Resolution: Eight bits in 250mS (10 bits in 1 second typical, >10 bits in >1 second
typical).
Update Rate: 4 per second
Ranges: 0-20mA, 4-20mA, 0-10V, and 0-5V*
Load Impedance: 0-20mA: 500 ohm maximum
4-20mA: 500 ohm maximum
0- 10V: 500 ohm minimum
0-5V: 500 ohm minimum
Isolation: Isolated from all other inputs and outputs.
* Changes between V and mA ranges also require jumper movement.