Setup Parameters
Note: When changing between °C and °F, the setup parameters Filt, o1.Hy, HSP.L, and HSP.H
settings convert automatically.
inPt Input type – this selects thermocouple type. LBB ovens use type J
thermocouple.
unit Process unit – selects between °C and °F for reading process
temperature.
rESo Display resolution – selects the location of the decimal point on process
related parameters.
SHif PV shift value (offset) – this moves the display temperature to the oven
temperature.
Filt PV filter – if process value is unstable to read, increasing this value will
steady the input signal.
out1 Output 1 function – this the function of the output. This must be set to .Hi
for
HI-LIMIT control.
o1.Hy Output 1 hysteresis value – the amount of degrees that the temperature
must be below the setpoint temperature before the HI-LIMIT can be reset.
HSP.L Lower limit of HSP1 – the minimum temperature that the HI-LIMIT can be
set.
HSP.H Upper limit of HSP1 – the maximum temperature that the HI-LIMIT can be
Code
Name
Settings
inPt Input type. J_tC
unit Process unit.
°
°°
°
C (or
°
°°
°
F)
rESo Display resolution. No.dP
SHif PV shift value (offset). 0
Filt PV filter. 0
out1 Output 1 function. Hi.
o1.Hy Output 1 hysteresis value. 2.0
HSP.L Lower limit of HSP1.
0 (32 if
°
°°
°
F)
HSP.H Upper limit of HSP1.
204 (400 if
°
°°
°
F)
out2 Output 2 function. (N/A) None
Addr Address assignment for digital communication. (N/A) 1
bAud Baud rate of digital communication. (N/A) 4.8
PAri Parity bit of digital communication. (N/A) EVEn
diSP Normal display format. PV